Related Experiment Videos
Percutaneous access devices in calves receiving an artificial heart
1Artificial Heart Laboratory, University of Utah, Salt Lake City.
Artificial Organs
|February 1, 1988
Summary
Infections from percutaneous access devices (PAD) are a major cause of death in calves with total artificial hearts (TAH). A new PAD design and implantation technique reduced superficial infections in calves with TAH.
Area of Science:
- Veterinary medicine
- Biomedical engineering
- Cardiovascular research
Background:
- Vegetative endocarditis and thromboembolism are primary causes of mortality in calves supported by total artificial hearts (TAH).
- Percutaneous access devices (PAD) for TAH drive lines are frequently infected, serving as a source for vegetative lesions.
- Infections often originate at the dorsal surface of the PAD, linked to skin ulceration or avulsion.
Purpose of the Study:
- To investigate the causes of PAD infections in calves with TAH.
- To evaluate the efficacy of a new PAD design and implantation technique in reducing infections.
Main Methods:
- Observation of 80 calves with TAHs to identify infection origins.
- Necropsy evaluation of 58 PAD specimens for flange positioning and infection.
- Comparative analysis of infection rates between old and new PAD designs in 11 animals.
Main Results:
- Improper flange positioning (76%) and gross infection (57%) were prevalent in older PAD designs.
- The new implantation technique and PAD design improved flange positioning and reduced skin interface irritation.
- Necropsies of animals with the new PAD showed a significant reduction in superficial infections.
Conclusions:
- PAD infections are a critical issue in TAH research in calves.
- Improved PAD design and implantation technique effectively mitigate superficial infections.