Fat-Water Phantoms for Magnetic Resonance Imaging Validation: A Flexible and Scalable Protocol
Published on: September 7, 2018
Temperature-corrected proton density fat fraction estimation using chemical shift-encoded MRI in phantoms
Ruvini Navaratna1,2, Ruiyang Zhao1,2, Timothy J Colgan2
1Department of Medical Physics, University of Wisconsin - Madison, Madison, Wisconsin, USA.
Purpose:
Chemical shift-encoded MRI (CSE-MRI) is well-established to quantify proton density fat fraction (PDFF) as a quantitative biomarker of hepatic steatosis. However, temperature is known to bias PDFF estimation in phantom studies. In this study, strategies were developed and evaluated to correct for the effects of temperature on PDFF estimation through simulations, temperature-controlled experiments, and a multi-center, multi-vendor phantom study.
Theory And Methods:
A technical solution that assumes and automatically estimates a uniform, global temperature throughout the phantom is proposed. Computer simulations modeled the effect of temperature on PDFF estimation using magnitude-, complex-, and hybrid-based CSE-MRI methods. Phantom experiments were performed to assess the temperature correction on PDFF estimation at controlled phantom temperatures. To assess the temperature correction method on a larger scale, the proposed method was applied to data acquired as part of a nine-site multi-vendor phantom study and compared to temperature-corrected PDFF estimation using an a priori guess for ambient room temperature.
Results:
Simulations and temperature-controlled experiments show that as temperature deviates further from the assumed temperature, PDFF bias increases. Using the proposed correction method and a reasonable a priori guess for ambient temperature, PDFF bias and variability were reduced using magnitude-based CSE-MRI, across MRI systems, field strengths, protocols, and varying phantom temperature. Complex and hybrid methods showed little PDFF bias and variability both before and after correction.
Conclusion:
Correction for temperature reduces temperature-related PDFF bias and variability in phantoms across MRI vendors, sites, field strengths, and protocols for magnitude-based CSE-MRI, even without a priori information about the temperature.
More Related Videos
Related Concept Videos
Proton (¹H) NMR: Chemical Shift
Absorption signals of all the protium nuclei...
¹³C NMR: Distortionless Enhancement by Polarization Transfer (DEPT)
NMR Spectroscopy: Chemical Shift Overview
For instance, the proton...
¹H NMR Chemical Shift Equivalence: Homotopic and Heterotopic Protons
¹H NMR Chemical Shift Equivalence: Enantiotopic and Diastereotopic Protons
In chiral compounds such as 2-butanol, replacing the methylene hydrogens at C3 produces a pair of...
Chemical Shift: Internal References and Solvent Effects
The internal reference compound generally used in NMR spectroscopy is tetramethylsilane (TMS). TMS is preferred because it is chemically inert, soluble in NMR solvents, and easily removable. Also, the highly shielded methyl protons in TMS yield an intense...