Related Experiment Video
Updated: Nov 18, 2025

A Detailed Protocol for Perspiration Monitoring Using a Novel, Small, Wireless Device
Published on: November 24, 2016
An Innovative Wearable Device For Monitoring Continuous Body Surface Temperature (HEARThermo): Instrument Validation
Chun-Yin Yeh1,2, Yi-Ting Chung2, Kun-Ta Chuang1
1Department of Computer Science and Information Engineering, National Cheng Kung University, Tainan, Taiwan.
Background:
Variations in body temperature are highly informative during an illness. To date, there are not many adequate studies that have investigated the feasibility of a wearable wrist device for the continuous monitoring of body surface temperatures in humans.
Objective:
The objective of this study was to validate the performance of HEARThermo, an innovative wearable device, which was developed to continuously monitor the body surface temperature in humans.
Methods:
We implemented a multi-method research design in this study, which included 2 validation studies-one in the laboratory and one with human subjects. In validation study I, we evaluated the test-retest reliability of HEARThermo in the laboratory to measure the temperature and to correct the values recorded by each HEARThermo by using linear regression models. We conducted validation study II on human subjects who wore HEARThermo for the measurement of their body surface temperatures. Additionally, we compared the HEARThermo temperature recordings with those recorded by the infrared skin thermometer simultaneously. We used intraclass correlation coefficients (ICCs) and Bland-Altman plots to analyze the criterion validity and agreement between the 2 measurement tools.
Results:
A total of 66 participants (age range, 10-77 years) were recruited, and 152,881 completed data were analyzed in this study. The 2 validation studies in the laboratory and on human skin indicated that HEARThermo showed a good test-retest reliability (ICC 0.96-0.98) and adequate criterion validity with the infrared skin thermometer at room temperatures of 20°C-27.9°C (ICC 0.72, P<.001). The corrected measurement bias averaged -0.02°C, which was calibrated using a water bath ranging in temperature from 16°C to 40°C. The values of each HEARThermo improved by the regression models were not significantly different from the temperature of the water bath (P=.19). Bland-Altman plots showed no visualized systematic bias. HEARThermo had a bias of 1.51°C with a 95% limit of agreement between -1.34°C and 4.35°C.
Conclusions:
The findings of our study show the validation of HEARThermo for the continuous monitoring of body surface temperatures in humans.
More Related Videos
Related Concept Videos
Assessing Body Temperature - Axilla
Step 1: Perform hand hygiene and put on clean gloves to maintain infection control and prevent cross-contamination.
Step 2: Prepare the patient by explaining the procedure to ensure understanding and cooperation. Ensure privacy, expose the axilla, and inform the patient that minimal movement is crucial for an accurate reading.
Step 3: Adjust the patient’s clothing to expose only the axilla. It minimizes...
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Assessing Body Temperature - Tympanic membrane
Step 1: Begin by practicing good hand hygiene to prevent the transmission of microorganisms.
Step 2: Turn on the thermometer and wait until the ready sign appears on the screen to ensure accurate measurement.
Step 3: Slide the probe cover in place to prevent cross-contamination.
Step 4: Instruct the patient to tilt their head to the side for comfort and check for cerumen...
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's...
Assessing Body Temperature - Rectal
Follow these steps for rectal temperature assessment:
Step 1: Perform hand hygiene and don clean gloves to prevent cross-infection.
Step 2: Position the patient in a side-lying position to better visualize the rectal...
Assessing Body Temperature - Oral
Step 1:
Start by practicing proper hand hygiene to prevent the spread of microorganisms.
Step 2:
Take the thermometer out of the charging unit, switch it on, and wait for the ready sign.
Step 3:
Gently slide the probe cover until a click is heard. This simple action prevents cross-contamination and ensures the correct placement of the probe cover.
Step 4:
Instruct the patient to open their mouth and place...