Comparison of optical performance among three dental operating microscopes: A pilot study
Poorya Jalali1, Charles Kim1, Karl F Woodmansey1
1Departments of Endodontics, Texas A and M College of Dentistry, Dallas, Texas, USA.
Introduction:
Two important aspects of the dental operating microscope (DOM) that factor into its overall effectiveness are resolution and depth of field. Therefore, the objective of this study was to evaluate and compare the resolution and depth of field of DOMs from three well-known manufacturers using standardized test targets.
Materials And Methods:
A resolution test, using the 1951 USAF Hi-Resolution Target (Edmund Optics, Barrington, NJ), and a depth of field test, using the Depth of Field Target 5-15 (Edmund Optics, Barrington, NJ), were performed by two calibrated observers. Three DOM systems such as Seiler IQ (Seiler Instrument Inc., St. Louis, USA), Global G-Series 6 step (Global Surgical Corp., St. Louis, USA), and Zeiss Extaro 300 (Carl Zeiss Meditec AG, Oberkochen, Germany) were used to compare the resolution and depth of field.
Results:
The Zeiss Extaro 300 showed the highest maximum resolution and maximum DOF (64 lp/mm and 17mm, respectively). The Seiler IQ showed the lowest maximum resolution and maximum DOF (35.9 lp/mm and 11 mm, respectively).
Conclusion:
Within the limitations of this study, the Zeiss Extaro 300 was superior in terms of resolution and depth of field as compared to the other two DOMs.
More Related Videos
Related Concept Videos
Imaging Biological Samples with Optical Microscopy
In optical microscopy, the specimen to be viewed is placed on a glass slide and clipped on the stage...
Three-Dimensional Microscopy in Microbiology
Confocal Fluorescence Microscopy
Overview of Microscopy Techniques
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
Two-Dimensional Microscopy in Microbiology