High-Sensitivity Micro LC-MS/MS Assay for Serum Estradiol without Derivatization.
Xin Yi1,2, Edward Ki Yun Leung1, Rachael Bridgman1
1Department of Pathology, The University of Chicago, Chicago, IL.
The Journal of Applied Laboratory Medicine
|February 25, 2021
Summary
This study presents a new micro LC-MS/MS method for measuring low estradiol (E2) levels without derivatization. The validated assay offers high sensitivity and a low limit of quantification, crucial for various clinical applications.
Area of Science:
- Biochemistry
- Analytical Chemistry
- Clinical Chemistry
Background:
- Accurate measurement of low estradiol (E2) concentrations (<20 pg/mL) is critical for postmenopausal women, men, pediatric patients, and breast cancer patients.
- Existing high-sensitivity LC-MS/MS methods for E2 often require large sample volumes and complex derivatization steps.
- There is a need for a simpler, more sensitive method to measure E2 at clinically relevant low levels.
Purpose of the Study:
- To develop and validate a high-sensitivity, underivatized micro LC-MS/MS method for measuring estradiol (E2).
- To achieve a reliable quantification of E2 concentrations below 5 pg/mL using a minimal sample volume.
Main Methods:
- Estradiol (E2) was extracted from a small sample volume (290 μL) using a hexane/ethyl acetate mixture.
- Separation was performed on a micro LC system (Eksigent Ekspert™) with a flow rate of 35 μL/min over 3.5 minutes.
- Detection was achieved using a SCIEX QTRAP 6500 mass spectrometer in negative mode, utilizing specific transitions for E2 and its deuterated internal standard (E2-d4).
Main Results:
- The method demonstrated a broad linear range (3.0-820.0 pg/mL) with excellent linearity (r2 > 0.999).
- The limit of quantification (LOQ) was determined to be 3.0 pg/mL, with total precision below 15% at all quality control levels.
- The assay showed good correlation with estradiol radioimmunoassay (RIA) (r2 = 0.96) and modest correlation with an automated immunoassay (r2 = 0.86).
Conclusions:
- A routinely applicable micro LC-MS/MS method for quantifying estradiol (E2) in blood samples was successfully developed and validated.
- The method is highly sensitive, achieving an LOQ of 3.0 pg/mL without the need for derivatization.
- This underivatized approach simplifies sample preparation and reduces sample volume requirements for E2 analysis.