Effect of implantation site of the His bundle pacing leads on pacing parameters: a single-center experience
Oushan Tang1, Haoliang Zhou2, Caidi Yuan3
1Department of Cardiology, The Second Hospital Affiliated to ShaoXing University, 123 Yan'an Road, Shaoxing, 312000, Zhejiang Province, China. tangoush@usx.edu.cn.
Background:
HB pacing is a promising approach to achieve physiological pacing, but its efficacy and long-term effects require further validation. In current study, we deemed to investigate the effect of the His bundle pacing (HBP) lead location on pacing parameters.
Methods:
2D echocardiography imaging was performed after successful implantation, according to which the patients were divided into groups A (whose His lead tips were at the atrial side) and B (whose His lead tips were at the ventricular side). The capture thresholds, sensing values, and H-V intervals between the two groups were compared.
Results:
Thirteen patients were in group A and 16 patients were in group B. The average capture thresholds during, 1 month, and 1 year after operation were 1.20 ± 0.34, 0.69 ± 0.29, and 0.92 ± 0.80 V/0.5 ms for group A and 1.14 ± 0.43, 0.81 ± 0.39, and 0.98 ± 0.59 V/0.5 ms for group B, respectively. The difference between the two groups was not significant. The threshold values in both groups decreased significantly in 1 month and slightly increased in 1 year. The sensing values of group A were 1.87 ± 0.82, 1.95 ± 0.76, and 1.88 ± 0.75 mV, while those of group B were 4.53 ± 1.37, 4.69 ± 1.38, and 4.59 ± 1.42 mV. The difference among the three time points was not significant. However, the sensing values in group A were consistently significantly lower than those in group B. The HV interval in group A was significantly longer than that in group B.
Conclusions:
The implantation site of HBP leads has a significant effect on sensing values for that His leads crossing the tricuspid annulus toward the ventricle are associated with higher sensing values, compared to a more proximal location. Meanwhile, lead location has no evident effect on capture thresholds that is improved significantly shortly after operation.
More Related Videos
12:45Benefits of Cardiac Resynchronization Therapy in an Asynchronous Heart Failure Model Induced by Left Bundle Branch Ablation and Rapid Pacing
Published on: December 11, 2017
10:17Real-Time Cardiac Mapping with a Noninvasive Imageless Electrocardiographic Imaging System
Published on: April 11, 2025
Related Concept Videos
Pulse Assessment Sites
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...