Infant heart-rate measurement and oxygen desaturation detection with a digital video camera using imaging
Matthew E Wieler1,2, Thomas G Murphy3,4, Mira Blecherman5
1Women & Infants' Hospital, The Warren Alpert Medical School of Brown University, Providence, RI, USA. matthew.wieler@gmail.com.
Objective:
To assess the feasibility of using an ordinary digital video camera to measure heart rate and detect oxygen desaturations in healthy infants.
Study Design:
Heart rate and oxygen saturation were measured with a video camera by detecting small color changes in 28 infants' foreheads and compared with standard pulse oximetry measures. Multivariable regression examined the relationship between infant characteristics and heart-rate measurement precision.
Results:
The average bias of camera heart-rate measures was -4.2 beats per minute (BPM) and 95% limits of agreement were ±43.8 BPM. Desaturations detected by camera were 75% sensitive (15/20) and had a positive predictive value of 20% (15/74). Lower birth-weight was independently correlated with more precise heart-rate measures (8.05 BPM per kg, [95% CI 0.764-15.3]).
Conclusions:
A digital video camera provides accurate but imprecise measures of infant heart rate and may provide a rough screening tool for oxygen desaturations.
More Related Videos
Related Concept Videos
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.