Retrospective Cardiac Gating with A Prototype Small-Animal X-ray Computed Tomograph
Published on: February 21, 2025
Adaptive CaRdiac cOne BEAm computed Tomography (ACROBEAT): Developing the next generation of cardiac cone beam CT
Tess Reynolds1, Owen Dillon1, Joseph Prinable1
1Faculty of Medicine and Health, ACRF Image X Institute, University of Sydney, Sydney, NSW, 2006, Australia.
Purpose:
An important factor when considering the use of interventional cone beam computed tomography (CBCT) imaging during cardiac procedures is the trade-off between imaging dose and image quality. Accordingly, Adaptive CaRdiac cOne BEAm computed Tomography (ACROBEAT) presents an alternative acquisition method, adapting the gantry velocity and projection rate of CBCT imaging systems in accordance with a patient's electrocardiogram (ECG) signal in real-time. The aim of this study was to experimentally investigate that ACROBEAT acquisitions deliver improved image quality compared to conventional cardiac CBCT imaging protocols with fewer projections acquired.
Methods:
The Siemens ARTIS pheno (Siemens Healthcare, GmbH, Germany), a robotic CBCT C-arm system, was used to compare ACROBEAT with a commercially available conventional cardiac imaging protocol that utilizes multisweep retrospective ECG-gated acquisition. For ACROBEAT, real-time control of the gantry position was enabled through the Siemens Test Automation Control system. ACROBEAT and conventional image acquisitions of the CIRS Dynamic Cardiac Phantom were performed, using five patient-measured ECG traces. The traces had average heart rates of 56, 64, 76, 86, and 100 bpm. The total number of acquired projections was compared between the ACROBEAT and conventional acquisition methods. The image quality was assessed via the contrast-to-noise ratio (CNR), structural similarity index (SSIM), and root-mean square error (RMSE).
Results:
Compared to the conventional protocol, ACROBEAT reduced the total number of projections acquired by 90%. The visual image quality provided by the ACROBEAT acquisitions, across all traces, matched or improved compared to conventional acquisition and was independent of the patient's heart rate. Across all traces, ACROBEAT averaged 1.4 times increase in the CNR, a 23% increase in the SSIM and a 29% decrease in the RMSE compared to conventional and was independent of the patient's heart rate.
Conclusion:
Adaptive patient imaging is feasible on a clinical robotic CBCT system, delivering higher quality images while reducing the number of projections acquired by 90% compared to conventional cardiac imaging protocols.
More Related Videos
09:32Time-Resolved, Dynamic Computed Tomography Angiography for Characterization of Aortic Endoleaks and Treatment Guidance via 2D-3D Fusion-Imaging
Published on: December 9, 2021
09:49Dual-phase Cone-beam Computed Tomography to See, Reach, and Treat Hepatocellular Carcinoma during Drug-eluting Beads Transarterial Chemo-embolization
Published on: December 2, 2013
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...