Agreement between continuous noninvasive finger cuff-derived and invasive arterial blood pressure measurements:
Moritz Flick1, Jasmin Matin Mehr, Luisa Briesenick
1From the Department of Anesthesiology, Center of Anesthesiology and Intensive Care Medicine, University Medical Center Hamburg-Eppendorf, Hamburg, Germany (MF, JMM, LB, PH, KK, CV, BS), CNSystems Medizintechnik, Graz, Austria (DF, KL), and Outcomes Research Consortium, Cleveland, Ohio, USA (BS).
Background:
The effect of different methods for data sampling and data processing on the results of comparative statistical analyses in method comparison studies of continuous arterial blood pressure (AP) monitoring systems remains unknown.
Objective:
We sought to investigate the effect of different methods for data sampling and data processing on the results of statistical analyses in method comparison studies of continuous AP monitoring systems.
Design:
Prospective observational study.
Setting:
University Medical Center Hamburg-Eppendorf, Hamburg, Germany, from April to October 2019.
Patients:
49 patients scheduled for neurosurgery with AP measurement using a radial artery catheter.
Main Outcome Measures:
We assessed the agreement between continuous noninvasive finger cuff-derived (CNAP Monitor 500; CNSystems Medizintechnik, Graz, Austria) and invasive AP measurements in a prospective method comparison study in patients having neurosurgery using all beat-to-beat AP measurements (Methodall), 10-s averages (Methodavg), one 30-min period of 10-s averages (Method30), Method30 with additional offset subtraction (Method30off), and 10 30-s periods without (Methodiso) or with (Methodiso-zero) application of the zero zone. The agreement was analysed using Bland-Altman and error grid analysis.
Results:
For mean AP, the mean of the differences (95% limits of agreement) was 9.0 (-12.9 to 30.9) mmHg for Methodall, 9.2 (-12.5 to 30.9) mmHg for Methodavg, 6.5 (-9.3 to 22.2) mmHg for Method30, 0.5 (-9.5 to 10.5) mmHg for Method30off, 4.9 (-6.0 to 15.7) mmHg for Methodiso, and 3.4 (-5.9 to 12.7) mmHg for Methodiso-zero. Similar trends were found for systolic and diastolic AP. Results of error grid analysis were also influenced by using different methods for data sampling and data processing.
Conclusion:
Data sampling and data processing substantially impact the results of comparative statistics in method comparison studies of continuous AP monitoring systems. Depending on the method used for data sampling and data processing, the performance of an AP test method may be considered clinically acceptable or unacceptable.
More Related Videos
07:35Simultaneous Electrocardiography Recording and Invasive Blood Pressure Measurement in Rats
Published on: January 31, 2019
10:21Continuous Blood Sampling in Small Animal Positron Emission Tomography/Computed Tomography Enables the Measurement of the Arterial Input Function
Published on: August 8, 2019
Related Concept Videos
Assessment of blood pressure in brachial artery(two-step method)
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Pre-Procedural Guidelines for Assessing Blood Pressure
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure: