Modelling of energy-dependent spectral resolution for SPECT Monte Carlo simulations using SIMIND
Michaella Morphis1, Johan A van Staden1, Hanlie du Raan1
1Department of Medical Physics, Faculty of Health Sciences, University of the Free State, Bloemfontein, South Africa.
Purpose:
Monte Carlo (MC) modelling techniques have been used extensively in Nuclear Medicine (NM). The theoretical energy resolution relationship ( ), does not accurately predict the gamma camera detector response across all energies. This study aimed to validate the accuracy of an energy resolution model for the SIMIND MC simulation code emulating the Siemens Symbia T16 dual-head gamma camera.
Methods:
Measured intrinsic energy resolution data (full width half maximum (FWHM) values), for Ba-133, Lu-177, Am-241, Ga-67, Tc-99m, I-123, I-131 and F-18 sources in air, were used to create a fitted model of the energy response of the gamma camera. Both the fitted and theoretical models were used to simulate intrinsic and extrinsic energy spectra using three different scenarios (source in air; source in simple scatter phantom and a clinical voxel-based digital patient phantom).
Results:
The results showed the theoretical model underestimated the FWHM values at energies above 160.0 keV up to 23.5 keV. In contrast, the fitted model better predicted the measured FWHM values with differences less than 3.3 keV. The I-131 in-scatter energy spectrum simulated with the fitted model better matched the measured energy spectrum. Higher energy photopeaks, (I-123: 528.9 keV and I-131: 636.9 keV) simulated with the fitted model, more accurately resembled the measured photopeaks. The voxel-based digital patient phantom energy spectra, simulated with the fitted and theoretical models, showed the potential impact of an incorrect energy resolution model when simulating isotopes with multiple photopeaks.
Conclusion:
Modelling of energy resolution with the proposed fitted model enables the SIMIND user to accurately simulate NM images. A great improvement was seen for high-energy photon emitting isotopes (e.g. I-131), as well as isotopes with multiple photopeaks (e.g. Lu-177, I-131 and Ga-67) in comparison to the theoretical model. This will result in accurate evaluation of radioactivity quantification, which is vital for dosimetric purposes.
Related Concept Videos
NMR Spectrometers: Resolution and Error Correction
Maxwell-Boltzmann Distribution: Problem Solving
This distribution function f(v) is defined by saying that the expected number N (v1,v2) of particles with speeds between v1 and v2 is given by
Mechanistic Models: Compartment Models in Algorithms for Numerical Problem Solving
In individual population analyses, different algorithms are employed, such as Cauchy's method, which uses a...
IR Spectroscopy: Hooke's Law Approximation of Molecular Vibration
According to Hooke's law, the vibrational frequency is directly proportional to...