Related Experiment Video
Updated: Nov 15, 2025

Upper-extremity Approach for Secondary Access in Transfemoral Transcatheter Aortic Valve Implantation
Published on: August 8, 2025
Snaring the Transcatheter Heart Valve Delivery System During Aortic Valve Replacement: When and Why
Andrea Buono1, Massimo Medda1, Sigitas Cesna2
1Clinical and Interventional Cardiology Unit, Istituto Clinico Sant'Ambrogio, Gruppo Ospedaliero San Donato, Milan, Italy.
Abstract:
None of the commercially available self-expanding (SE)-trans-catheter heart valve (THVs) used for trans-catheter aortic valve replacement (TAVR) has a deflectable delivery system. This aspect can represent a shortcoming, especially in some anatomical scenario where THV advancement is challenging. The use of a snare catheter (SC), applying an external traction force, can increase THV trackability. We describe three different TAVR cases, in which a SE-THV was the only available option and where the THV delivery system advancement in the aortic arch was not feasible, if not with the combined use of a SC.
Related Concept Videos
Cardiac Catheterization III: Left Heart Catheterization
Cardiac Catheterization IV: Nursing Management
Cardiac Catheterization I: Pre-Procedure Overview