Does the beta regularization parameter of bayesian penalized likelihood reconstruction always affect the
Zhifang Wu1,2, Binwei Guo1, Bin Huang1
1Department of Nuclear Medicine, First Hospital of Shanxi Medical University, Taiyuan, Shanxi, P.R. China.
Purpose:
This study aims to provide a detailed investigation on the noise penalization factor in Bayesian penalized likelihood (BPL)-based algorithm, with the utilization of partial volume effect correction (PVC), so as to offer the suitable beta value and optimum standardized uptake value (SUV) parameters in clinical practice for small pulmonary nodules.
Methods:
A National Electrical Manufacturers Association (NEMA) image-quality phantom was scanned and images were reconstructed using BPL with beta values ranged from 100 to 1000. The recovery coefficient (RC), contrast recovery (CR), and background variability (BV) were measured to assess the quantification accuracy and image quality. In the clinical assessment, lesions were categorized into sub-centimeter (<10 mm, n = 7) group and medium size (10-30 mm, n = 16) group. Signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) were measured to evaluate the image quality and lesion detectability. With PVC was performed, the impact of beta values on SUVs (SUVmax, SUVmean, SUVpeak) of small pulmonary nodules was evaluated. Subjective image analysis was performed by two experienced readers.
Results:
With the increasing of beta values, RC, CR, and BV decreased gradually in the phantom work. In the clinical study, SNR and CNR of both groups increased with the beta values (P < 0.001), although the sub-centimeter group showed increases after the beta value reached over 700. In addition, highly significant negative correlations were observed between SUVs and beta values for both lesion-size groups before the PVC (P < 0.001 for all). After the PVC, SUVpeak measured from the sub-centimeter group was no significantly different among different beta values (P = 0.830).
Conclusion:
Our study suggests using SUVpeak as the quantification parameter with PVC performed to mitigate the effects of beta regularization. Beta values between 300 and 400 were preferred for pulmonary nodules smaller than 30 mm.
More Related Videos
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
06:33Construction of a Preclinical Multimodality Phantom Using Tissue-mimicking Materials for Quality Assurance in Tumor Size Measurement
Published on: July 29, 2013
Related Concept Videos
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...