Cuff-less blood pressure estimation from photoplethysmography signal and electrocardiogram
Li-Ping Yao1,2, Zhong-Liang Pan3
1College of Physics and Telecommunications Engineering, South China Normal University, Guangzhou, 510006, China.
Abstract:
In recent studies, the physiological parameters derived from human vital signals are found as the status response of the heart and arteries. In this paper, we therefore firstly attempt to extract abundant vital features from photoplethysmography(PPG) signal, its multivariate derivative signals and Electrocardiogram(ECG) signal, which are verified its statistical significance in BP estimation through statistical analysis t-test. Afterwards, the optimal feature set are obtained by usnig mutual information coefficient analysis, which could investigate the potential associations with blood pressure. The optimized feature set are aid as an input to various machine learning strategies for BP estimation. The results indicates that AdaBoost based BP estimation model outperforms other regression methods. Concurrently, AdaBoost-based model is further analyzed by using the Histograms of Estimation Error and Bland-Altman Plot. The results also indicate the great BP estimation performance of the proposed BP estimation method, and it stays within the Advancement of Medical Instrumention(AAMI) standard. Regarding the British Hypertension Society (BHS), it achieves the grade A for DBP and grade B for MAP. Besides, the experimental result illustrated that our proposed BP estimation method could reduce the MAE and the STD, and improve the r for SBP, MAP and DBP estimation, respectively, which further demonstrates the feasibility of our proposed BP estimation method in this paper.
More Related Videos
Related Concept Videos
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and...
Measurement of Blood Pressure