Cross-Calibration of Prodigy and Horizon A Densitometers and Precision of the Horizon A Densitometer
Elise Reitshamer1, Kelsey Barrett1, Kyla Shea1
1Jean Mayer USDA Human Nutrition Research Center on Aging, Tufts University, Boston, MA, USA.
Abstract:
We performed this study to enable a reliable transition for clinical study participants and patients from a GE Lunar Prodigy to a Hologic Horizon A dual-energy X-ray absorptiometry (DXA) scanner and to assess the reproducibility of measurements made on the new DXA scanner. Forty-five older adults had one spine, hip, and total body scan on a Prodigy dual-energy X-ray absorptiometry (DXA) scanner and 2 spine, hip, and total body scans, with repositioning, on a new Hologic Horizon A DXA scanner. Linear regression models were used to derive cross calibration equations for each measure on the 2 scanners. Precision (group root-mean-square average coefficient of variation) of bone mineral density (BMD) of the total hip, femoral neck, and lumbar spine (L1-L4), and total body fat, bone, and lean mass, appendicular lean mass, and trabecular bone score (TBS) was assessed using the International Society of Clinical Densitometry's (ISCD's) Advanced Precision Calculation Tool. Correlation coefficients for the BMD and body composition measures on the 2 scanners ranged from 0.94 to 0.99 (p<0.001). When compared with values on the Prodigy, mean BMD on the Horizon A was lower at each skeletal site (0.136 g/cm2 lower at the femoral neck and 0.169 g/cm2 lower at the lumbar spine (L1-4)), fat mass was 0.47 kg lower, and lean mass was 4.50 kg higher. Precision of the Horizon A scans was 1.60% for total hip, 1.94% for femoral neck, and 1.25% for spine (L1-4) BMD. Precision of TBS was 1.67%. Precision of total body fat mass was 2.16%, total body lean mass was 1.26%, appendicular lean mass was 1.97%, and total body bone mass was 1.12%. The differences in BMD and body composition values on the 2 scanners illustrate the importance of cross-calibration to account for these differences when transitioning clinical study participants and patients from one scanner to another.
More Related Videos
09:10Precision of In Vivo Quantitative Tooth Wear Measurement Using Intra-Oral Scans
Published on: July 12, 2022
14:11Quantification of Hydrogen Concentrations in Surface and Interface Layers and Bulk Materials through Depth Profiling with Nuclear Reaction Analysis
Published on: March 29, 2016
Related Concept Videos
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Calibration Curves: Correlation Coefficient
Uncertainty in Measurement: Reading Instruments
Difference from Background: Limit of Detection
The LOD indicates the presence or absence...
Glassware Calibration
Volumetric flasks: Volumetric flasks are designed to prepare aqueous solutions of precise volumes accurately with a calibration line on the neck. To calibrate a volumetric flask, it is important to fill it with distilled...
Distance Corrections