Related Experiment Video
Updated: Nov 12, 2025

A Novel Method to Determine the Longitudinal Antibacterial Activity of Drug-Eluting Materials
Published on: March 3, 2023
Sensitization to Implant Components Is Associated with Joint Replacement Failure: Identification and Revision to
Annyce S Mayer1, Samantha Erb2, Raymond H Kim3
1Division of Environmental and Occupational Health Sciences, Department of Medicine, National Jewish Health, Denver, Colo; Division of Environmental and Occupational Health, University of Colorado School of Medicine, School of Public Health, Aurora, Colo.
Background:
Over 90% of one million annual US joint replacements are highly successful. Nonetheless, 10% do poorly owing to infection or mechanical issues. Many implant components are sensitizers, and sensitization could also contribute to implant failure.
Objective:
To determine the prevalence of implant sensitization in joint failure patients, their clinical characteristics, and implant revision outcomes. We hypothesized that sensitized patients would improve when revised with nonallergenic materials.
Methods:
We prospectively enrolled 105 joint failure patients referred by orthopedic surgeons who had already excluded infection or mechanical causes. Patients provided informed consent, completed a history and physical examination, patch testing to metals and bone cement, and a nickel lymphocyte proliferation test. A study coordinator was able to contact 64% of patients (n = 67) 9 to 12 months later to evaluate outcomes.
Results:
A total of 59% were sensitized to an implant component: 32% to metal and 37% to bone cement. The nickel lymphocyte proliferation test was 60% sensitive and 96% specific in diagnosing nickel sensitization. Most sensitized subjects reported no or uncertain histories of reactions to a specific material. Implant sensitized patients were younger and reported previous eczema, joint itching, and implant loosening. By 9 to 12 months later, most patients with a revised implant (revised) described significant improvement (16 of 22 revised for sensitization [P = .0003] vs 9 of 13 revised without sensitization [P = .047]) compared with patients without implant revision). All revised patients with sensitization used components to which they were not sensitized. Pain (P = .001), swelling (P = .035), and instability (P = .006) were significantly reduced in the revised sensitized group.
Conclusions:
Sensitization to implant components is an important cause of unexplained joint replacement failure. Joint revisions based on sensitization information resulted in significant improvements.
Related Concept Videos
Allergic Drug Reactions
Cross-reactivity
Hypersensitivities
Types of Hypersensitivities
Hypersensitivity reactions are categorized into four types: Type 1, Type 2, Type 3, and Type 4. Each type has a distinct mechanism...
Allergic Reactions
Cell-mediated Immune Responses
Tissue Transplantation
The Biology of Tissue Transplantation
The biology of tissue transplantation hinges on the Major Histocompatibility Complex (MHC) molecules. These molecules...