Related Experiment Video
Updated: Nov 12, 2025

Treatment with Locking Intramedullary Nailing for Intertrochanteric Fracture of the Femur Utilizing a New Awl with a Distal Positioner
Published on: June 6, 2025
Gross Trunnion Failure in the Bipolar Hemiarthroplasty; Raising Concern about Short Trunnion: A Case Report
Kangbaek Kim1, Junpyo Lee1, Jeong Joon Yoo1
1Department of Orthopaedic Surgery, Seoul National University Hospital, Seoul, Korea.
Abstract:
There have been some reports of gross trunnion failure (GTF) in total hip arthroplasty. Here, we report a case of GTF 19 years after bipolar hemiarthroplasty using a 28-mm head with a 14/16 taper bore. Compared to other GTF reports, the current case had some unusual aspects: bipolar hemiarthroplasty, 28-mm head, relatively late-onset, and no apparent findings of metallosis though a severe one was evident. A Computed tomography scout view provided valuable information in evaluating polyethylene, metal head, and neck inside the bipolar cup. The current report suggests a need for concerned regarding short trunnion length which may be associated with GTF.