Wideband Optical Detector of Ultrasound for Medical Imaging Applications
Published on: May 11, 2014
Sub-Nanowatt Ultrasonic Bio-Telemetry Using B-Scan Imaging
Sri Harsha Kondapalli1, Shantanu Chakrabartty1
1Department of Electrical and Systems Engineering at Washington University in St. Louis, St. Louis, MO 63130 USA.
Goal:
The objective of this paper is to investigate if the use of a B-scan ultrasound imaging system can reduce the energy requirements, and hence the power-dissipation requirements to support wireless bio-telemetry at an implantable device.
Methods:
B-scan imaging data were acquired using a commercial 256-element linear ultrasound transducer array which was driven by a commercial echoscope. As a transmission medium, we used a water-bath and the operation of the implantable device was emulated using a commercial-off-the-shelf micro-controller board. The telemetry parameters (e.g. transmission rate and transmission power) were wirelessly controlled using a two-way radio-frequency transceiver. B-scan imaging data were post-processed using a maximum-threshold decoder and the quality of the ultrasonic telemetry link was quantified in terms of its bit-error-rate (BER).
Results:
Measured results show that a reliable B-scan communication link with an implantable device can be achieved at transmission power levels of 100 pW and for implantation depths greater than 10 cm.
Conclusions:
In this paper we demonstrated that a combination of B-scan imaging and a simple decoding algorithm can significantly reduce the energy-budget requirements for reliable ultrasonic telemetry.