Multiplexed measurement of candidate blood protein biomarkers of heart failure
Claire Tonry1, Ken McDonald2, Mark Ledwidge2
1Wellcome-Wolfson Institute for Experimental Medicine, Queen's University Belfast, 97 Lisburn Rd, Belfast, BT9 7BL, UK.
Aims:
There is a critical need for better biomarkers so that heart failure can be diagnosed at an earlier stage and with greater accuracy. The purpose of this study was to design a robust mass spectrometry (MS)-based assay for the simultaneous measurement of a panel of 35 candidate protein biomarkers of heart failure, in blood. The overall aim was to evaluate the potential clinical utility of this biomarker panel for prediction of heart failure in a cohort of 500 patients.
Methods And Results:
Multiple reaction monitoring (MRM) MS assays were designed with Skyline and Spectrum Mill PeptideSelector software and developed using nanoflow reverse phase C18 chromatographic Chip Cube-based separation, coupled to a 6460 triple quadrupole mass spectrometer. Optimized MRM assays were applied, in a sample-blinded manner, to serum samples from a cohort of 500 patients with heart failure and non-heart failure (non-HF) controls who had cardiovascular risk factors. Both heart failure with reduced ejection fraction (HFrEF) patients and heart failure with preserved ejection fraction (HFpEF) patients were included in the study. Peptides for the Apolipoprotein AI (APOA1) protein were the most significantly differentially expressed between non-HF and heart failure patients (P = 0.013 and P = 0.046). Four proteins were significantly differentially expressed between non-HF and the specific subtypes of HF (HFrEF and HFpEF); Leucine-rich-alpha-2-glycoprotein (LRG1, P < 0.001), zinc-alpha-2-glycoprotein (P = 0.005), serum paraoxanse/arylesterase (P = 0.013), and APOA1 (P = 0.038). A statistical model found that combined measurements of the candidate biomarkers in addition to BNP were capable of correctly predicting heart failure with 83.17% accuracy and an area under the curve (AUC) of 0.90. This was a notable improvement on predictive capacity of BNP measurements alone, which achieved 77.1% accuracy and an AUC of 0.86 (P = 0.005). The protein peptides for LRG1, which contributed most significantly to model performance, were significantly associated with future new onset HF in the non-HF cohort [Peptide 1: odds ratio (OR) 2.345 95% confidence interval (CI) (1.456-3.775) P = 0.000; peptide 2: OR 2.264 95% CI (1.422-3.605), P = 0.001].
Conclusions:
This study has highlighted a number of promising candidate biomarkers for (i) diagnosis of heart failure and subtypes of heart failure and (ii) prediction of future new onset heart failure in patients with cardiovascular risk factors. Furthermore, this study demonstrates that multiplexed measurement of a combined biomarker signature that includes BNP is a more accurate predictor of heart failure than BNP alone.
More Related Videos
07:08A High Throughput, Multiplexed and Targeted Proteomic CSF Assay to Quantify Neurodegenerative Biomarkers and Apolipoprotein E Isoforms Status
Published on: October 20, 2016
08:50Multiplexed Fluorescent Microarray for Human Salivary Protein Analysis Using Polymer Microspheres and Fiber-optic Bundles
Published on: October 10, 2013
Related Concept Videos
Blood Studies for Cardiovascular System II: CRP, Hcy, and Cardiac Natriuretic Peptide Markers
These markers indicate stress or strain on the heart muscle:
Natriuretic Peptides (BNP)
Cardiac myocytes produce these hormones in response to ventricular stretching...
Blood Studies for Cardiovascular System I: Cardiac Biomarkers
The essential diagnostic tools for detecting myocardial necrosis and monitoring individuals suspected of having acute coronary syndrome (ACS) include:
Troponins
Troponins, particularly cardiac troponins I and T, are the most precise and sensitive markers of myocardial injury. They are detectable within 4-6 hours of myocardial injury and remain...