Performance evaluation of bactericidal effect and endotoxin inactivation by low-temperature ozone/hydrogen peroxide
Yusuke Nomura1, Junji Yamamura2, Chie Fukui1
1Division of Medical Devices, National Institute of Health Sciences, Kawasaki-shi, Kanagawa, Japan.
Abstract:
This study evaluated the performance of a new O3 /H2 O2 mixed gas sterilization instrument for killing microorganisms and inactivating bacterial endotoxin at low temperatures. Sterility assurance level was achieved by an over 6-log reduction of Geobacillus stearothermophilus ATCC 12980, and the decimal reduction value was 0.77 min in sterilization mode. A reduction of over 3 logs in Limulus amebocyte lysate coagulation activity of purified endotoxin from Escherichia coli was observed after treatment in endotoxin-inactivation mode. The same inactivation ability was observed when treating dried bacterial cells. Biomaterials made of polymer or metal did not exhibit cytotoxicity after gas exposure at O3 concentrations below 200 ppm. As the results of human cell-based pyrogen testing, significant amounts of endotoxin that were over the limit for medical devices contacting cerebrospinal fluid (2.15 EU/device) were detected on scissors washed with a washer-disinfector and sterilized with ethylene oxide or autoclaving. In contrast, endotoxin decreased to 0.29 ± 0.05 EU/device after O3 /H2 O2 mixed gas sterilization in endotoxin-inactivation mode. Compared to conventional gas sterilization methods, O3 /H2 O2 mixed gas has high sterilization ability and a strong capacity to inactivate endotoxin. It is expected that this sterilization technology will improve the safety of reusable medical devices and utensils for regenerative medicine.
More Related Videos
Related Concept Videos
Physical Methods for Controlling Microbial Growth: Temperature
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Methods of Sterilization I: Physical Methods
Steam sterilization uses non-toxic, low-cost moist heat in the form of saturated steam under pressure, which is fast, microbicidal, and sporicidal, and quickly warms and penetrates fabrics. Autoclaves, or steam sterilizers, expose each item to direct steam contact for a predetermined time at the necessary...
Physical Methods for Controlling Microbial Growth: Radiation and Filtration
Antimicrobial Effectiveness
Oxidative Cleavage of Alkenes: Ozonolysis
Ozone is a symmetrical bent molecule stabilized by a resonance structure.