Microstructural and Corrosion Properties of Hydroxyapatite Containing PEO Coating Produced on AZ31 Mg Alloy
Luca Pezzato1, Katya Brunelli1, Stefano Diodati2,3
1Department of Industrial Engineering, University of Padua, Via Marzolo 9, 35131 Padova, Italy.
Abstract:
In this work, the composition of an electrolyte was selected and optimized to induce the formation of hydroxyapatite during Plasma electrolytic oxidation (PEO) treatment on an AZ31 alloy for application in bioabsorbable implants. In detail, the PEO process, called PEO-BIO (Plasma Electrolytic Oxidation-Biocompatible), was performed using a silicate-phosphate-based electrolyte with the addition of calcium oxide in direct-current mode using high current densities and short treatment times. For comparison, a known PEO process for producing anticorrosive coatings, called standard, was applied on the same alloy. The coatings were characterized by scanning electron microscopy (SEM), X-ray diffraction (XRD) and XPS analyses. The corrosion performance was evaluated in simulated body fluid (SBF) at 37 °C. The coating produced on the PEO-BIO sample was porous and thicker than the standard PEO one, with zones enriched in Ca and P. The XRD analysis showed the formation of hydroxyapatite and calcium oxides in addition to magnesium-silicon oxide and magnesium oxide in the PEO-BIO sample. The corrosion resistance of PEO-BIO sample was comparable with that of a traditional PEO treated sample, and higher than that of the untreated alloy.
More Related Videos
07:14Synthesis of Graphene-Hydroxyapatite Nanocomposites for Potential Use in Bone Tissue Engineering
Published on: July 27, 2022
12:18Co-localizing Kelvin Probe Force Microscopy with Other Microscopies and Spectroscopies: Selected Applications in Corrosion Characterization of Alloys
Published on: June 27, 2022