Related Experiment Video
Updated: Nov 10, 2025

Differential Imaging of Biological Structures with Doubly-resonant Coherent Anti-stokes Raman Scattering CARS
Published on: October 17, 2010
A Short Review of Cavity-Enhanced Raman Spectroscopy for Gas Analysis
Christian Niklas1, Hainer Wackerbarth1, Georgios Ctistis1
1Institut für Nanophotonik Göttingen e.V., Hans-Adolf-Krebs-Weg 1, 37077 Göttingen, Germany.
Abstract:
The market of gas sensors is mainly governed by electrochemical, semiconductor, and non-dispersive infrared absorption (NDIR)-based optical sensors. Despite offering a wide range of detectable gases, unknown gas mixtures can be challenging to these sensor types, as appropriate combinations of sensors need to be chosen beforehand, also reducing cross-talk between them. As an optical alternative, Raman spectroscopy can be used, as, in principle, no prior knowledge is needed, covering nearly all gas compounds. Yet, it has the disadvantage of a low quantum yield through a low scattering cross section for gases. There have been various efforts to circumvent this issue by enhancing the Raman yield through different methods. For gases, in particular, cavity-enhanced Raman spectroscopy shows promising results. Here, cavities can be used to enhance the laser beam power, allowing higher laser beam-analyte interaction lengths, while also providing the opportunity to utilize lower cost equipment. In this work, we review cavity-enhanced Raman spectroscopy, particularly the general research interest into this topic, common setups, and already achieved resolutions.
More Related Videos
09:46Direct Comparison of Hyperspectral Stimulated Raman Scattering and Coherent Anti-Stokes Raman Scattering Microscopy for Chemical Imaging
Published on: April 28, 2022
07:52A Novel Technique for Raman Analysis of Highly Radioactive Samples Using Any Standard Micro-Raman Spectrometer
Published on: April 12, 2017
Related Concept Videos
Raman Spectroscopy Instrumentation: Overview
The monochromatic laser source, typically using visible or near-infrared radiation, generates a highly focused beam of light. This light interacts with the molecules of the sample, scattering some of the light. Liquid and gaseous samples are usually tested in ordinary glass capillaries, while solids can be analyzed as powders packed in capillaries or as potassium...
Raman Spectroscopy: Overview
However, a small fraction of the scattered light exhibits a frequency shift due to the exchange of energy between the incident photons and...
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall....
Gas Chromatography: Types of Detectors-II
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...