Automated Midline Shift and Intracranial Pressure Estimation based on Brain CT Images
Published on: April 13, 2013
Imaging pediatric acute head trauma using 100-kVp low dose CT with adaptive statistical iterative reconstruction
Yanan Zhu1, Zhian Pi1, Heping Zhou1
1Medical Imaging Centre, Affiliated Hospital of Ankang University (Ankang Central Hospital), Ankang, China.
Objective:
To demonstrate the ability of achieving low dose and high-quality head CT images for children with acute head trauma using 100 kVp and adaptive statistical iterative reconstruction (ASIR-V) algorithm in single rotation on a 16 cm wide-detector system.
Materials And Methods:
We retrospectively analyzed the CT dose index (CTDI) and image quality of 104 children aged 0-6 years with acute head trauma (1 hour -3 days) in two groups: Group 1(n = 50) on a 256-row CT with single rotation at a reduced-dose of 100 kVp/240 mA and reconstructed using ASIR-V at 70%level; Group 2(n = 54) on a 64-row CT with multiple rotations at a standard dose of 120 kVp/ 180mA and reconstructed using a conventional filtered back-projection (FBP). Both groups used the 0.5 s/r axial scan mode. CT dose index (CTDI) and quantitative image quality measurements were compared using the Student t test; qualitative image quality comparison was carried out using Mann-Whitney rank test and the inter-reviewer agreement was evaluated using Kappa test.
Results:
The exposure time was 0.5 s for Group 1 and 3.27±0.29 s for Group 2. The CTDI in Group 1 was 9.74±0.86mGy, 36.38%lower than the 15.31mGy in Group 2 (p < 0.001). Group 1 and Group 2 had similar artifact index (2.06±1.06 vs. 2.37±1.18) in the cerebellar hemispheres, and similar contrast-to-noise ratio (2.32±0.83 vs. 1.69±0.68), (1.47±0.72 vs. 1.10±0.43) respectively for cerebellum and thalamus (p > 0.05). Image quality was acceptable for diagnosis, and motion artifacts were reduced in Group 1 (p < 0.001).
Conclusion:
Single rotation CT with 100 kVp and 70%ASIR-V on 16 cm wide-detector CT reduces radiation dose and motion artifacts for children with acute head trauma without compromising diagnostic quality as compared with standard dose protocol. Thus, it provides a novel imaging method in management of pediatric acute head trauma.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...
Imaging Studies for Cardiovascular System V: CT