[About a case of macro-FSH analytical interference]
Claire Guibet1, Cindy Lauro2, Virginie Grouthier3
1Service de médecine nucléaire, CHU de Bordeaux, Bordeaux, France.
Annales De Biologie Clinique
|April 5, 2021
Summary
Elevated follicle-stimulating hormone (FSH) levels in an 18-year-old woman were caused by a macro-FSH complex, not premature ovarian failure. This finding highlights the importance of investigating analytical interferences in hormone testing.
Area of Science:
- Endocrinology
- Reproductive Medicine
- Clinical Chemistry
Background:
- Premature ovarian failure (POF) is a concern in reproductive-aged women presenting with menstrual irregularities.
- Elevated follicle-stimulating hormone (FSH) with normal luteinizing hormone (LH) levels can suggest POF, necessitating thorough investigation.
Observation:
- An 18-year-old woman exhibited menstrual cycle disorders and elevated FSH with normal LH, initially suspected as POF.
- Initial investigations suggested analytical interference rather than true ovarian dysfunction.
Findings:
- The elevated FSH levels were ultimately attributed to a macro-FSH complex, formed by FSH bound to an autoantibody.
- Heterophilic antibodies and human anti-mouse antibodies (HAMA) were considered but ruled out as the primary cause.
Implications:
- This case underscores the critical need to consider analytical interferences, such as macro-FSH, in diagnosing endocrine disorders.
- Accurate interpretation of gonadotropin levels requires awareness of potential artifactual elevations to avoid misdiagnosis of premature ovarian failure.
Related Concept Videos
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
821
Inductively coupled plasma–mass spectrometry (ICP–MS) is a highly selective and sensitive technique for accurate elemental analysis. Though the analysis of ICP–MS mass spectra is comparatively straightforward, it is affected by spectroscopic and non-spectroscopic interferences. Spectroscopic interferences arise when the plasma contains ionic species with an m/z value the same as the analyte ion. Spectroscopic interference can be categorized as isobaric, polyatomic ions, and...
821
Atomic Absorption Spectroscopy: Interference
1.6K
Interference leads to systematic error in atomic absorption (AA) measurements by enhancing or diminishing the analytical signal or the background. These interferences can be grouped into three main categories: spectral interference, chemical interference, and physical interference.
Spectral interference occurs when signals from other elements or molecules overlap with the analyte signal, falsely elevating or masking the analyte's absorbance. This interference can be corrected using Zeeman,...
Spectral interference occurs when signals from other elements or molecules overlap with the analyte signal, falsely elevating or masking the analyte's absorbance. This interference can be corrected using Zeeman,...
1.6K