Unexplained syncope during pacemaker interrogation: Wand, AutoCapture, or lead-Which one is the elusive culprit?
Debabrata Bera1, Suchit Majumder2, Subir Ghose1
1Department of Cardiology, RTIICS, Kolkata, West Bengal, India.
Abstract:
A 65-year-old gentleman underwent dual chamber pacemaker implantation (DDDR, St Jude Medical) 7 years back for infra-hisian complete heart block. He was completely asymptomatic and came for his annual routine check-up. After undergoing ECG with and without magnet, he was prepared for device evaluation. After placing the programmer wand over the chest as soon as the ";interrogate" button on the programmer screen was pressed, the patient immediately experienced pre-syncope but recovered instantly as the wand was promptly withdrawn. After taking him to the casualty room with all resuscitation measures in hand, a repeat attempt of interrogation was made after connecting ECG, which revealed reproducible loss of capture (LOC), exclusively during wand placement. A differential diagnosis of lead failure, battery depletion, or wand related issues were considered. However, serial ECGs recorded without wand raised the possibility of AutoCapture malfunction. With all precautions, the device was programmed to fixed ventricular output mode after which interrogation could be performed safely. There was a remaining battery longevity of 2 years with acceptable lead parameters and stable threshold. He continues to be asymptomatic at 10 months of follow up.
Related Concept Videos
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias
Dysrhythmias IV: Characteristics of Bradyarrhythmias
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and...
Electrophysiology of Normal Cardiac Rhythm
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin...