RF shielding materials for highly-integrated PET/MRI systems
Laiyin Yin1, Franziska Schrank1, Nicolas Gross-Weege1
1Department of Physics of Molecular Imaging Systems, Institute for Experimental Molecular Imaging, RWTH Aachen University, Aachen, Germany.
Researchers evaluated radio frequency (RF) shielding materials for simultaneous positron emission tomography (PET) and magnetic resonance imaging (MRI) systems. Coated foils like HS 9400 offer a good balance of shielding effectiveness and minimal impact on PET and MRI performance.
Area of Science:
- Medical Imaging Physics
- Materials Science for Medical Devices
- Radio Frequency Engineering
Background:
- Simultaneous PET/MRI systems require novel radio frequency (RF) shielding solutions to mitigate interference.
- Conventional PET detector housings limit integration possibilities; excluding components like scintillators offers advantages but necessitates RF shielding interfaces.
- Selecting appropriate RF shielding materials is critical for maintaining both PET and MRI performance.
Purpose of the Study:
- To evaluate the shielding effectiveness (SE) and impact on positron emission tomography (PET) performance of various RF shielding materials.
- To assess the magnetic resonance imaging (MRI) compatibility of these materials by analyzing their effect on magnetic field homogeneity.
- To identify optimal RF shielding materials for integrated PET/MRI systems.
Main Methods:
- Examined 10 different RF shielding materials, including coated glasses, coated foils, and meshes.
- Measured shielding effectiveness (SE) at 100 and 300 MHz.
- Assessed PET performance by measuring photon attenuation (PA), energy resolution (dE/E), and coincidence resolving time (CRT).
- Evaluated MRI compatibility by analyzing the impact on B0 field homogeneity.
Main Results:
- Coated glasses and foils achieved SEs up to 25 dB at 300 MHz with low photon attenuation (<23%) and minimal impact on PET performance (dE/E, CRT) and B0 homogeneity.
- Meshes offered higher SEs (up to 56 dB) but resulted in significantly higher photon attenuation (>58%) and greater degradation of PET performance metrics.
- Coated foil HS 9400 demonstrated excellent performance with SE = 25 dB, PA = 22%, dE/E = 12%, and CRT = 274 ps, with no measurable impact on B0 homogeneity.
Conclusions:
- Coated foils and glasses are promising candidates for RF shielding interfaces in integrated PET/MRI systems due to their balance of shielding capability and minimal performance impact.
- Meshes provide superior shielding but their high photon attenuation and impact on PET performance limit their suitability for this application.
- The coated foil HS 9400 is recommended for PET/MRI integration concepts, offering a robust solution for effective RF shielding without compromising image quality or system compatibility.