Diagnostic Accuracy of Ambulatory Devices in Detecting Atrial Fibrillation: Systematic Review and Meta-analysis
Tien Yun Yang1, Li Huang2, Shwetambara Malwade3
1School of Medicine, College of Medicine, Taipei Medical University, Taipei, Taiwan.
Insights
Ambulatory devices using non-12-lead electrocardiograms (ECG) and photoplethysmography (PPG) demonstrate high accuracy for detecting atrial fibrillation (AF). These technologies show promise as convenient screening tools for early AF diagnosis.
Area of Science:
- Cardiology
- Medical Devices
- Diagnostic Accuracy
Background:
- Atrial fibrillation (AF) is a prevalent cardiac arrhythmia requiring early diagnosis to mitigate severe health consequences.
- Traditional 12-lead ECG is the standard for AF diagnosis, but its application can be limited.
- Ambulatory devices offer potential for convenient AF screening.
Purpose of the Study:
- To evaluate the diagnostic accuracy of non-12-lead ECG and photoplethysmography (PPG) in ambulatory devices for AF detection.
- To compare the accuracy of these novel technologies against the gold standard 12-lead ECG.
- To analyze the influence of study design and participant risk on diagnostic performance.
Main Methods:
- A systematic review and meta-analysis adhering to PRISMA guidelines.
- Searched MEDLINE and EMBASE databases for studies from 2015 to 2021.
- Employed a bivariate model to pool sensitivity, specificity, likelihood ratios, and SROC; assessed study quality using QUADAS-2.
Main Results:
- Included 16 studies with 3217 participants; analyzed 7623 assessments.
- Non-12-lead ECG showed pooled sensitivity of 89.7% and specificity of 95.7%.
- PPG demonstrated higher pooled sensitivity (94.7%) and specificity (97.6%) for AF detection.
Conclusions:
- Both non-12-lead ECG and PPG exhibit high diagnostic accuracy for AF.
- These ambulatory technologies can function as effective preliminary screening tools.
- Further validation in diverse populations is recommended due to observed variations in high-risk vs. low-risk groups.
Background:
Atrial fibrillation (AF) is the most common cardiac arrhythmia worldwide. Early diagnosis of AF is crucial for preventing AF-related morbidity, mortality, and economic burden, yet the detection of the disease remains challenging. The 12-lead electrocardiogram (ECG) is the gold standard for the diagnosis of AF. Because of technological advances, ambulatory devices may serve as convenient screening tools for AF.
Objective:
The objective of this review was to investigate the diagnostic accuracy of 2 relatively new technologies used in ambulatory devices, non-12-lead ECG and photoplethysmography (PPG), in detecting AF. We performed a meta-analysis to evaluate the diagnostic accuracy of non-12-lead ECG and PPG compared to the reference standard, 12-lead ECG. We also conducted a subgroup analysis to assess the impact of study design and participant recruitment on diagnostic accuracy.
Methods:
This systematic review and meta-analysis was conducted in accordance with the Preferred Reporting Items for Systematic Reviews and Meta-Analysis (PRISMA) guidelines. MEDLINE and EMBASE were systematically searched for articles published from January 1, 2015 to January 23, 2021. A bivariate model was used to pool estimates of sensitivity, specificity, positive likelihood ratio (PLR), negative likelihood ratio (NLR), and area under the summary receiver operating curve (SROC) as the main diagnostic measures. Study quality was evaluated using the quality assessment of diagnostic accuracy studies (QUADAS-2) tool.
Results:
Our search resulted in 16 studies using either non-12-lead ECG or PPG for detecting AF, comprising 3217 participants and 7623 assessments. The pooled estimates of sensitivity, specificity, PLR, NLR, and diagnostic odds ratio for the detection of AF were 89.7% (95% CI 83.2%-93.9%), 95.7% (95% CI 92.0%-97.7%), 20.64 (95% CI 10.10-42.15), 0.11 (95% CI 0.06-0.19), and 224.75 (95% CI 70.10-720.56), respectively, for the automatic interpretation of non-12-lead ECG measurements and 94.7% (95% CI 93.3%-95.8%), 97.6% (95% CI 94.5%-99.0%), 35.51 (95% CI 18.19-69.31), 0.05 (95% CI 0.04-0.07), and 730.79 (95% CI 309.33-1726.49), respectively, for the automatic interpretation of PPG measurements.
Conclusions:
Both non-12-lead ECG and PPG offered high diagnostic accuracies for AF. Detection employing automatic analysis techniques may serve as a useful preliminary screening tool before administering a gold standard test, which generally requires competent physician analyses. Subgroup analysis indicated variations of sensitivity and specificity between studies that recruited low-risk and high-risk populations, warranting future validity tests in the general population.
Trial Registration:
PROSPERO International Prospective Register of Systematic Reviews CRD42020179937; https://www.crd.york.ac.uk/prospero/display_record.php?RecordID=179937.
More Related Videos
Related Concept Videos
Dysrhythmias V: Evaluating Dysrhythmias
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Holter Monitor: 24-Hour Monitoring
Acute Coronary Syndrome III: Diagnostic Studies