Performance of the RT-LAMP-based eazyplex® SARS-CoV-2 as a novel rapid diagnostic test
Renate Egerer1, Birgit Edel1, Bettina Löffler1
1Institute of Medical Microbiology, Jena University Hospital, Jena, Germany.
Background:
Diagnostic assays for severe acute respiratory syndrome Coronavirus-2 (SARS-CoV-2) that are easy to perform and produce fast results are essential for timely decision making regarding the isolation of contagious individuals.
Objective:
We evaluated the CE-approved eazyplex® SARS-CoV-2, a ready-to-use real time RT-LAMP assay for identification of the SARS-CoV-2 N and ORF8 genes from swabs in less than 30 min without RNA extraction.
Study Design:
Oropharyngeal and nasal swabs from 100 positive and 50 negative patients were inoculated into 0.9 % saline and tested by NeuMoDx™ RT-PCR. An aliquot was diluted fivefold in Copan sputum liquefying (SL) solution and directly analyzed by eazyplex® SARS-CoV-2. In addition, 130 patient swabs were prospectively tested with both methods in parallel. Analytical sensitivity of the assay was determined using virus stock dilutions.
Results:
Positive percent agreement (PPA) between the eazyplex® SARS-CoV-2 and RT-PCR was 74 % for samples with Ct values < 35. When using a Ct cut-off ≤ 28 the PPA increased to 97.4 %. In the prospective part of the study overall PPA of the eazyplex® kit was 66.7 % but increased to 100 % when only Ct values ≤ 28 were considered. There were no false positive results. The median time to positivity was 12.5 min for the N gene and 16.75 min for ORF8. Analytical sensitivity was 3.75 TCID50/mL. 105 virus copies/mL were reproducibly detected.
Conclusion:
The eazyplex® SARS-CoV-2 is a rapid assay that accurately identifies samples with high viral loads. It may be useful for near-patient testing outside of a molecular diagnostic laboratory.