Related Experiment Video
Updated: Sep 28, 2026

Large Scale Non-targeted Metabolomic Profiling of Serum by Ultra Performance Liquid Chromatography-Mass Spectrometry UPLC-MS
Published on: March 14, 2013
The analytical performance of six urine drug screens on cobas 6000 and ARCHITECT i2000 compared to LC-MS/MS gold
Mohamed Abou El Hassan1, David Colantonio2, Sergei Likhodii3
1Clinical Chemistry Division, Provincial Laboratory Services, Queen Elizabeth Hospital, Charlottetown, Prince Edward Island, Canada; Department of Pathology, Dalhousie University, Halifax, NS, Canada.
Background:
Immunoassays provide a rapid tool for the screening of drugs-of-abuse (DOA). However, results are presumptive and confirmatory testing is warranted. To reduce associated cost and delay, laboratories should employ assays with high positive and negative predictive values (PPVs and NPVs). Here, we compared the results of urine drug screens on cobas 6000 (cobas) and ARCHITECTi2000 (ARCHITECT) platforms for six drugs against LC-MS/MS to assess the analytical performance of these assays.
Methods:
Eighty nine residual urine specimens, which tested positive for amphetamine, THC-COOH, benzoylecgonine, EDDP, opiates and/or oxycodone during routine drug testing, were stored frozen until later confirmation by LC-MS/MS. Immunoassays were performed on cobas and ARCHITECT using a split sample. A third aliquot from these samples was tested by LC-MS/MS to assess the percentage of false positive, false negative, true positive and true negative results and calculate the PPVs and NPVs for each immunoassay.
Results:
The PPVs of THC-COOH and EDDP assays were 100% on both platforms. Suboptimal PPVs were achieved for oxycodone (cobas, 57.1% vs ARCHITECT, 66.7%), amphetamine (77.8 vs. 100%), opiates (80.0 vs. 84.6%) and benzoylecgonine (88.9 vs. 84.2%) assays. The NPV was 100% for cobas and ARCHITECT oxycodone assays. Lower NPVs were achieved for THC-COOH (cobas, 28.6% vs ARCHITECT, 25.0%), EDDP (72.7% for both assays), benzoylecgonine (74.4% vs 73.8%), amphetamine (83.3% vs 82.8%) and opiates (100% vs 85.3%).
Conclusion:
Overall, cobas and ARCHITECT urine drug screens have comparable analytical performance. Confirmatory testing is warranted for positive test results especially for oxycodone, amphetamine, opiates and cocaine. Negative drug screen results must be interpreted with caution especially for THC-COOH, EDDP, benzoylecgonine, amphetamine and opiates.
More Related Videos
10:17High-throughput and Comprehensive Drug Surveillance Using Multisegment Injection-Capillary Electrophoresis-Mass Spectrometry
Published on: April 23, 2019
10:13Author Spotlight: An Efficient Methodology to Confidently Differentiate and Characterize Fentanyl Analogs
Published on: November 8, 2024