Double-barreled and Concentric Microelectrodes for Measurement of Extracellular Ion Signals in Brain Tissue
Published on: September 5, 2015
Three-micrometer-diameter needle electrode with an amplifier for extracellular in vivo recordings
Yuto Kita1, Shuhei Tsuruhara1, Hiroshi Kubo1
1Department of Electrical and Electric Information Engineering, Toyohashi University of Technology, 1-1 Hibarigaoka Tempaku-cho, 441-8580 Toyohashi, Japan.
Abstract:
Microscale needle-electrode devices offer neuronal signal recording capability in brain tissue; however, using needles of smaller geometry to minimize tissue damage causes degradation of electrical properties, including high electrical impedance and low signal-to-noise ratio (SNR) recording. We overcome these limitations using a device assembly technique that uses a single needle-topped amplifier package, called STACK, within a device of ∼1 × 1 mm2 Based on silicon (Si) growth technology, a <3-µm-tip-diameter, 400-µm-length needle electrode was fabricated on a Si block as the module. The high electrical impedance characteristics of the needle electrode were improved by stacking it on the other module of the amplifier. The STACK device exhibited a voltage gain of >0.98 (-0.175 dB), enabling recording of the local field potential and action potentials from the mouse brain in vivo with an improved SNR of 6.2. Additionally, the device allowed us to use a Bluetooth module to demonstrate wireless recording of these neuronal signals; the chronic experiment was also conducted using STACK-implanted mice.