An estimate of 90Y dosimetry for bremsstrahlung SPECT/CT imaging in liver therapy with 90Y microspheres
1Division of Nuclear Medicine, Department of Radiology, Icahn School of Medicine at Mount Sinai, Box 1141, One Gustave L. Levy Place, New York, NY, 10029, USA.
Objective:
Bremsstrahlung SPECT/CT (bSPECT/CT) is one of the most common methods for post-therapy imaging in 90Y microspheres selective internal radiation therapy (SIRT) of liver cancers. Here, we are proposing a simple approach using bSPECT/CT to estimate mean absorbed dose to the liver in patients undergoing treatment with 90Y microspheres.
Materials And Methods:
In our previous study comparing 90Y dosimetry obtained using bSPECT/CT vs PET/CT, we found that there was a large difference between the mean absorbed dose values to the whole-liver. However, there was a high linear correlation between the doses, which presented an opportunity for quantitative assessment using bSPECT/CT 90Y imaging. In this study, after treatment with 90Y microspheres, 43 patients were immediately imaged on a dual-head Infinia SPECT/CT gamma camera and on a mCT PET/CT system. Images from 25 of the patients, randomly selected, were used to calculate the correlation of mean liver doses obtained from bSPECT/CT vs. PET/CT. For the remaining 18 patients, the calculated correlation was used to estimate doses obtained from bSPECT/CT, and these estimations were then compared to the doses obtained from PET/CT, considered the gold standard for quantitative analysis.
Results:
From the 25 selected patients, the calculated linear correlation between bSPECT/CT and PET/CT 90Y mean absorbed doses in whole liver was high (r^2 = 0.97), with a slope of 2.80 and an intercept of -0.63. This linear fit was used to calculate the bSPECT/CT doses for the remaining 18 patients. For these patients, the mean whole-liver dose obtained from bSPECT/CT fitted data vs that obtained from PET/CT were 50.59 Gy and 50.81 Gy, respectively. The average dose difference was 0.2 ± 5.4 Gy (range -18.2%-13.0%). The repeatability coefficient was 10.5 (20.8 % of the mean).
Conclusion:
Although quantitative bremsstrahlung imaging is difficult, it is possible to calculate adequate estimates of whole-liver dosimetry from bSPECT/CT imaging that is calibrated using its correlation with post-therapy PET/CT 90Y images.