Heater-Cooler Devices and Risk of Contamination during Cardiac Surgery
Mirko Kaluza1,2, Gloria Färber1, Benjamin May1
1Department of Cardiothoracic Surgery, Jena University Hospital, Friedrich-Schiller University of Jena, Jena, Germany.
Background:
Heater-cooler devices (HCD) have been implicated in a cardiosurgical contamination scenario causing prosthetic valve endocarditis.
Aim:
We characterized contamination of new HCDs and assessed the risk of intraoperative microorganism transmission from the HCD to the operating field.
Methods:
We initially acquired four new FlexTherm and then four new Maquet HCU40 HCDs and assessed occurrence and speed of microbial contamination (including mycobacteria) assessing swab and water samples from the device. In parallel, we collected repeated samples from different sites in the operating room either by swab sticks or by exposing different sample plates to room air. We also reviewed microbiological results from the hospital and compared them to cardiosurgical wound infections and endocarditis cases. Finally, we simulated cardiosurgical conditions and assessed the devices' ability to expel air to the operative field.
Results:
All new HCDs were clean before first use. Despite authority-mandated decontamination procedures, microbial growth (Fusarium solani, Sphingomonas paucimobilis, Pseudomonas aeruginosa, Mycobacterium chelonae, and gordonae) was identified in all HCDs over time and could not be permanently eliminated. Four of these mircoorganisms were also found in tap water. However, none of the HCD-organisms were found inside the laminar airflow operating area. Importantly, except for P. aeruginosa, none of the HCD organisms were found in patients with surgical wound infections or endocarditis. HCD-expelled air did not rise more than 40 cm above ground.
Conclusion:
HCDs cannot be expected to remain sterile despite extensive decontamination procedures. However, airborne transmission of microorganisms directly from the HCD to the operating field appears unlikely.
Related Concept Videos
Cardiac Catheterization I: Pre-Procedure Overview
Cardiac Catheterization IV: Nursing Management
Cardiac Catheterization II: Right Heart Catheterization
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
Imaging Studies for Cardiovascular System V: CT
Cardiomyopathy VII: Pre and Post Operative Nursing Management