Comparison of Agreement and Accuracy using Binocular Wavefront Optometer with Autorefractor and Phoropter
Published on: September 16, 2025
Agreement and inter-session repeatability of manual and automatic interpupillary distance measurements
Liat Gantz1, Einat Shneor1, Ravid Doron1
1Department of Optometry and Vision Science, Hadassah Academic College, 37 Haneviim St, Jerusalem, 9101001 Israel.
Background:
Interpupillary distance (IPD) is important in developmental anatomy, genetics, design of optical instrumentation, ocular diagnostics, and optical prescribing. IPD frequently is measured on different days, and by either automatic pupillometers (physiological measurement) or manual ruler (anatomical measurement). Therefore, there is importance in the agreement and inter-session repeatability of manual and automatic IPD measurements.
Methods:
Monocular distance from the bridge of the nose and binocular distance and near binocular IPD were randomly measured, using a millimeter ruler and the Essilor Pupillon pupillometer. Gender effects were assessed using Wilcoxon and Mann-Whitney tests, respectively. Agreement was assessed using Spearman correlation and Bland-Altman (B&A) plots. Thirty additional participants were tested within 1-2 weeks to determine the inter-session repeatability.
Results:
The agreement study included 199 participants (mean age: 24.1 ± 5.0 range: 19-53, 58 male, 141 female) and the repeatability sub- study included 30 (mean age: 27.9 ± 4.5, range: 23-39, 6 male, 24 female). Males and females significantly differed in age (<2 year mean difference (md)) and IPD (monocular md: < 1 mm, binocular md: < 2 mm). Manual vs. automatic measurements were significantly different for all conditions (md: <1 mm for all) except for distance left eye male PD. There was no significant difference between the session for both methods.
Conclusions:
Binocular and monocular manual and automatic measurements were significantly different statistically, but not clinically. Distance binocular IPD was approximately 3 mm wider than near IPD. Male binocular IPD was approximately 2 mm wider than the female IPD. Both methods had good inter-session repeatability.