Validation of the TOPAS Monte Carlo toolkit for HDR brachytherapy simulations
Francisco Berumen1, Yunzhi Ma1, José Ramos-Méndez2
1Département de Radio-Oncologie et Axe oncologie du Centre de recherche du CHU de Québec, CHU de Québec, Québec, QC, Canada; Département de Physique, de Génie Physique et d'Optique et Centre de Recherche sur le Cancer, Université Laval, Québec, QC, Canada.
Purpose:
The goal of this work is to validate the user-friendly Geant4-based Monte Carlo toolkit TOol for PArticle Simulation (TOPAS) for brachytherapy applications.
Methods And Materials:
Brachytherapy simulations performed with TOPAS were systematically compared with published TG-186 reference data. The photon emission energy spectrum, the air-kerma strength, and the dose-rate constant of the model-based dose calculation algorithm (MBDCA)-WG generic Ir-192 source were extracted. For dose calculations, a track-length estimator was implemented. The four Joint AAPM/ESTRO/ABG MBDCA-WG test cases were evaluated through histograms of the local and global dose difference volumes. A prostate, a palliative lung, and a breast case were simulated. For each case, the dose ratio map, the histogram of the global dose difference volume, and cumulative dose-volume histograms were calculated.
Results:
The air-kerma strength was (9.772 ± 0.001) × 10-8 U Bq-1 (within 0.3% of the reference value). The dose-rate constant was 1.1107 ± 0.0005 cGy h-1 U-1 (within 0.01% of the reference value). For all cases, at least 96.9% of voxels had a local dose difference within [-1%, 1%] and at least 99.9% of voxels had a global dose difference within [-0.1%, 0.1%]. The implemented track-length estimator scorer was more efficient than the default analog dose scorer by a factor of 237. For all clinical cases, at least 97.5% of voxels had a global dose difference within [-1%, 1%]. Dose-volume histograms were consistent with the reference data.
Conclusions:
TOPAS was validated for high-dose-rate brachytherapy simulations following the TG-186 recommended approach for MBDCAs. Built on top of Geant4, TOPAS provides broad access to a state-of-the-art Monte Carlo code for brachytherapy simulations.
More Related Videos
08:25Radiation Planning Assistant - A Streamlined, Fully Automated Radiotherapy Treatment Planning System
Published on: April 11, 2018
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
Related Concept Videos
Radiation: Applications
The average...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...