Comparison between Widefield Optical Coherence Tomography Devices in Eyes with High Myopia
Federico Corvi1,2,3, Federico Zicarelli1, Matteo Airaldi1
1Eye Clinic, Department of Biomedical and Clinical Science "Luigi Sacco", Sacco Hospital, University of Milan, via G.B Grassi 74, 20157 Milan, Italy.
Four optical coherence tomography (OCT) devices were compared for imaging highly myopic eyes. Significant differences were found in retinal visualization, artifact prevalence, and retrobulbar tissue imaging among the OCT devices.
Area of Science:
- Ophthalmology
- Medical Imaging
- Retinal Imaging
Background:
- High myopia presents challenges for visualizing retinal and subretinal layers.
- Optical coherence tomography (OCT) is crucial for assessing ocular structures.
Purpose of the Study:
- To compare the performance of four OCT devices in imaging highly myopic eyes.
- To evaluate differences in retinal visualization, artifact prevalence, and retrobulbar tissue imaging.
Main Methods:
- Prospective, observational, cross-sectional study involving patients with high myopia and controls.
- Four OCT devices (Spectralis OCT2, PlexElite 2.0 100 kHz, PlexElite 2.0 200 kHz, Canon Xephilio OCT-S1) were used.
- Standardized line scans centered on the fovea were acquired for comparison.
Main Results:
- Canon Xephilio OCT-S1 visualized the largest retinal extent (p < 0.0001).
- Spectralis OCT showed a higher prevalence of conjugate image artifacts (p < 0.0001), while PlexElite 200 kHz showed a lower prevalence (p < 0.0001).
- Swept-source PlexElite devices visualized retrobulbar tissue in a higher proportion of eyes (p < 0.007).
Conclusions:
- OCT device choice impacts the visualization of retinal structures and retrobulbar tissue in high myopia.
- Significant differences exist in artifact prevalence and imaging extent among the tested OCT devices.
- These findings aid in selecting appropriate OCT technology for high myopia research and clinical practice.