Related Concept Videos

Positron Emission Tomography01:29

Positron Emission Tomography

6.5K
Positron emission tomography (PET) is a medical imaging technique involving radiopharmaceuticals — substances that emit short-lived radiation. Although the first PET scanner was introduced in 1961, it took 15 more years before radiopharmaceuticals were combined with the technique and revolutionized its potential.
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
6.5K
Imaging Studies II: Positron Emission Tomography and Scintigraphy01:25

Imaging Studies II: Positron Emission Tomography and Scintigraphy

298
Positron Emission Tomography (PET) is a medical imaging technique that provides crucial insights into the body's physiological functions at a molecular level. It is an indispensable resource for diagnosing, staging, and monitoring various illnesses, notably cancer, neurological disorders, and cardiovascular conditions.
Fundamental Principles of PET
298

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Evaluation of a PET Insert for Trimodal Imaging: A Step Toward PET/MRI-Guided Focused Ultrasound.

IEEE transactions on radiation and plasma medical sciences·2026
Same author

Magneto-stimulation limits in medical imaging applications with rapid field dynamics.

Physics in medicine and biology·2022
Same author

Architecture-Level Optimization on Digital Silicon Photomultipliers for Medical Imaging.

Sensors (Basel, Switzerland)·2022
Same author

Timing evaluation of a PET detector block based on semi-monolithic LYSO crystals.

Medical physics·2021
Same author

Fast Energy Dependent Scatter Correction for List-Mode PET Data.

Journal of imaging·2021
Same author

NEMA Performance Evaluation of CareMiBrain dedicated brain PET and Comparison with the whole-body and dedicated brain PET systems.

Scientific reports·2019

Related Experiment Video

Updated: Nov 7, 2025

Management of Respiratory Motion Artefacts in 18F-fluorodeoxyglucose Positron Emission Tomography using an Amplitude-Based Optimal Respiratory Gating Algorithm
06:53

Management of Respiratory Motion Artefacts in 18F-fluorodeoxyglucose Positron Emission Tomography using an Amplitude-Based Optimal Respiratory Gating Algorithm

Published on: July 23, 2020

5.8K

Simulation Study of a Frame-Based Motion Correction Algorithm for Positron Emission Imaging.

Héctor Espinós-Morató1, David Cascales-Picó1, Marina Vergara1,2

  • 1Instituto de Instrumentación para Imagen Molecular (i3M), Centro Mixto CSIC-Universitat Politècnica de València, 46022 Valencia, Spain.

Sensors (Basel, Switzerland)
|April 30, 2021
PubMed
Summary

This study introduces a novel frame-based algorithm to correct patient motion during Positron Emission Tomography (PET) scans. The method enhances image quality and accuracy, crucial for reliable diagnosis with advanced PET imaging.

Keywords:
motion correctionmulti-framesimulation studyspatio-temporal registration

More Related Videos

Creating Dynamic Images of Short-lived Dopamine Fluctuations with lp-ntPET: Dopamine Movies of Cigarette Smoking
14:21

Creating Dynamic Images of Short-lived Dopamine Fluctuations with lp-ntPET: Dopamine Movies of Cigarette Smoking

Published on: August 6, 2013

18.5K
Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
07:57

Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform

Published on: March 24, 2022

3.0K

Related Experiment Videos

Last Updated: Nov 7, 2025

Management of Respiratory Motion Artefacts in 18F-fluorodeoxyglucose Positron Emission Tomography using an Amplitude-Based Optimal Respiratory Gating Algorithm
06:53

Management of Respiratory Motion Artefacts in 18F-fluorodeoxyglucose Positron Emission Tomography using an Amplitude-Based Optimal Respiratory Gating Algorithm

Published on: July 23, 2020

5.8K
Creating Dynamic Images of Short-lived Dopamine Fluctuations with lp-ntPET: Dopamine Movies of Cigarette Smoking
14:21

Creating Dynamic Images of Short-lived Dopamine Fluctuations with lp-ntPET: Dopamine Movies of Cigarette Smoking

Published on: August 6, 2013

18.5K
Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
07:57

Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform

Published on: March 24, 2022

3.0K

Area of Science:

  • Medical Imaging
  • Nuclear Medicine
  • Image Processing

Background:

  • Positron Emission Tomography (PET) provides functional, non-invasive imaging using radiotracers to assess metabolic processes.
  • Advancements in PET scanner technology have improved spatial resolution but exacerbated image degradation due to involuntary patient motion.
  • Motion artifacts in PET imaging compromise diagnostic accuracy, necessitating effective motion correction techniques.

Purpose of the Study:

  • To develop and evaluate a novel frame-based algorithm for correcting motion artifacts in Positron Emission Tomography (PET) imaging.
  • To assess the algorithm's performance in improving image quality and diagnostic reliability in simulated PET data.
  • To investigate the algorithm's effectiveness in scenarios with limited statistical information and prior knowledge of lesions.

Main Methods:

  • A simulation study employing a frame-based algorithm for image motion correction in PET.
  • Data acquisition is segmented into frames, considering object size, to manage low statistical information.
  • Multi-level, spatio-temporal registration is utilized to align frames, followed by validation using simulated moving phantoms.

Main Results:

  • The proposed method effectively minimizes intra-frame motion, significantly improving signal intensity compared to existing techniques.
  • High similarity metrics were achieved between corrected images and ground-truth static images.
  • The algorithm demonstrated the capability to determine patient-injected dose limits when prior lesion knowledge is available.

Conclusions:

  • The developed frame-based motion correction algorithm enhances PET image quality and diagnostic accuracy.
  • This technique is particularly valuable for advanced PET scanners where motion artifacts are more pronounced.
  • The method offers a robust solution for motion correction, even with limited statistical data and potential prior lesion information.