Fabrication and Characterization of Optical Tissue Phantoms Containing Macrostructure
Published on: February 12, 2018
A calibration CT mini-lung-phantom created by 3-D printing and subtractive manufacturing
H Henry Guo1, Mats Persson2, Oliver Weinheimer3
1Department of Radiology, Stanford Medical Center, Stanford, CA, USA.
Abstract:
We describe the creation and characterization of a calibration CT mini-lung-phantom incorporating simulated airways and ground-glass densities. Ten duplicate mini-lung-phantoms with Three-Dimensional (3-D) printed tubes simulating airways and gradated density polyurethane foam blocks were designed and built. Dimensional accuracy and CT numbers were measured using micro-CT and clinical CT scanners. Micro-CT images of airway tubes demonstrated an average dimensional variation of 0.038 mm from nominal values. The five different densities of incorporated foam blocks, simulating ground-glass, showed mean CT numbers (±standard deviation) of -897.0 ± 1.5, -844.1 ± 1.5, -774.1 ± 2.6, -695.3 ± 1.6, and -351.0 ± 3.7 HU, respectively. Three-Dimensional printing and subtractive manufacturing enabled rapid, cost-effective production of ground-truth calibration mini-lung-phantoms with low inter-sample variation that can be scanned simultaneously with the patient undergoing lung quantitative CT.