Automated control of oxygen titration in preterm infants on non-invasive respiratory support
Peter A Dargaville1,2, Andrew P Marshall3, Oliver J Ladlow4
1Menzies Institute for Medical Research, University of Tasmania College of Health and Medicine, Hobart, Tasmania, Australia peter.dargaville@dhhs.tas.gov.au.
Objective:
To evaluate the performance of a rapidly responsive adaptive algorithm (VDL1.1) for automated oxygen control in preterm infants with respiratory insufficiency.
Design:
Interventional cross-over study of a 24-hour period of automated oxygen control compared with aggregated data from two flanking periods of manual control (12 hours each).
Setting:
Neonatal intensive care unit.
Participants:
Preterm infants receiving non-invasive respiratory support and supplemental oxygen; median birth gestation 27 weeks (IQR 26-28) and postnatal age 17 (12-23) days.
Intervention:
Automated oxygen titration with the VDL1.1 algorithm, with the incoming SpO2 signal derived from a standard oximetry probe, and the computed inspired oxygen concentration (FiO2) adjustments actuated by a motorised blender. The desired SpO2 range was 90%-94%, with bedside clinicians able to make corrective manual FiO2 adjustments at all times.
Main Outcome Measures:
Target range (TR) time (SpO2 90%-94% or 90%-100% if in air), periods of SpO2 deviation, number of manual FiO2 adjustments and oxygen requirement were compared between automated and manual control periods.
Results:
In 60 cross-over studies in 35 infants, automated oxygen titration resulted in greater TR time (manual 58 (51-64)% vs automated 81 (72-85)%, p<0.001), less time at both extremes of oxygenation and considerably fewer prolonged hypoxaemic and hyperoxaemic episodes. The algorithm functioned effectively in every infant. Manual FiO2 adjustments were infrequent during automated control (0.11 adjustments/hour), and oxygen requirements were similar (manual 28 (25-32)% and automated 26 (24-32)%, p=0.13).
Conclusion:
The VDL1.1 algorithm was safe and effective in SpO2 targeting in preterm infants on non-invasive respiratory support.
Trial Registration Number:
ACTRN12616000300471.
More Related Videos
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
07:52Expired CO2 Measurement in Intubated or Spontaneously Breathing Patients from the Emergency Department
Published on: January 29, 2011
Related Concept Videos
Physiological Control of Respiration
Breathing, a seemingly passive process, is regulated by the respiratory center in the brainstem. This center coordinates the involuntary control of respirations, which means it occurs without conscious effort, ensuring a smooth and uninterrupted pattern.
Regulation of Ventilation
The body maintains ventilation by monitoring levels of carbon dioxide (CO2), oxygen (O2), and hydrogen ion concentration (pH) in the arterial blood. Among these factors, the level of CO2 plays a crucial...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Respiratory Assessment: Purpose and Indications
Objectives and Importance:
The primary goal of respiratory assessment is to evaluate patients at early risk of clinical deterioration. Since respiratory distress often precedes other signs of declining health, breathing patterns and sounds become a...
Administering Oxygen by Mask
Administering oxygen by mask is a common nursing intervention that provides supplemental oxygen to patients with respiratory distress or chronic lung conditions. This procedure involves delivering oxygen at a specified rate through a face mask connected to an oxygen source.
Equipment
The equipment necessary for this procedure includes:
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Neural Control of Respiration
Respiratory Centers in the Brainstem
Two primary areas comprise the respiratory center: the medullary respiratory center in the medulla oblongata and the pontine respiratory group in the pons. The...