Not Removing the Glossy White Cover from Adhesive INVOS Neonatal Sensors Affects the Oxygenation Measurement
Mathias Lühr Hansen1, Daniel Ostojic2, Stefan Kleiser3
1Department of Neonatology, Rigshospitalet, University of Copenhagen, Copenhagen, Denmark.
Abstract:
The randomized clinical trial, SafeBoosC III, evaluates the effect of treatment guided by cerebral tissue oximetry monitoring in extremely preterm infants. Treatment should be considered, when cerebral oxygen saturation (StO2) drops below a predefined hypoxic threshold. This threshold value differs between different brands of instruments. To achieve high external validity, in this pragmatic trial all commercially available cerebral tissue oximeters have been accepted, provided their specific hypoxic threshold value has been determined in phantom studies. Since most companies produce sensors with an adhesive surface on the patient-contacting side, in the phantom studies these sensors were applied according to the specifications, i.e., the glossy cover was removed from the sensor. However, since the skin of preterm infants is particularly fragile, some neonatologists keep this cover on the adhesive sensors, to avoid the risk of skin injury when removing the sensor. Therefore, the aim of this study was to determine whether keeping this cover on leads to different StO2 values. To evaluate the effect of the cover, we performed multiple deoxygenations in a blood-lipid phantom and compared an INVOS neonatal sensor (Medtronic), with and without the cover, to a reference oximeter (OxiplexTS, ISS). As expected, the relationship of the StO2 between the INVOS neonatal sensor and OxiplexTS was linear (r2 = 0.999) with and without cover, but the cover influenced the linear equation: StO2_INVOS_cover = 1.133*StO2_ISS + 7.1 as opposed to StO2_INVOS_nocover = 1.103*StO2_ISS + 12.0. Furthermore, the hypoxic SafeBoosC III threshold differed as well: 60.3% with cover and 63.8% without cover. In conclusion, keeping the adhesive cover on an INVOS neonatal sensor results in lower measured values. At the hypoxic threshold, this is more than 3% (from 60.3% to 63.8%), and therefore, if clinicians keep the cover on the sensor, they need to be aware of this difference.
Related Concept Videos
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Administering Oxygen by Mask
Administering oxygen by mask is a common nursing intervention that provides supplemental oxygen to patients with respiratory distress or chronic lung conditions. This procedure involves delivering oxygen at a specified rate through a face mask connected to an oxygen source.
Equipment
The equipment necessary for this procedure includes:
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...
Administering Oxygen by Nasal Cannula
Nasal Cannulas
A nasal cannula is a lightweight tube split into two prongs placed in the nostrils,...