Comparing CSF amyloid-beta biomarker ratios for two automated immunoassays, Elecsys and Lumipulse, with amyloid PET
Eline A J Willemse1, Betty M Tijms2, Bart N M van Berckel3
1Department of Clinical Chemistry Neurochemistry Laboratory Amsterdam Neuroscience Vrije Universiteit Amsterdam Amsterdam UMC Amsterdam the Netherlands.
Introduction:
We evaluated for two novel automated biomarker assays how cerebrospinal fluid (CSF) amyloid beta (Aβ)1- 42-ratios improved the concordance with amyloid positron emission tomography (PET) positivity compared to Aβ1- 42 alone.
Methods:
We selected 288 individuals from the Amsterdam Dementia Cohort across the Alzheimer's disease clinical spectrum when they had both CSF and amyloid PET visual read available, regardless of diagnosis. CSF Aβ1- 42, phosphorylated tau (p-tau), and total tau (t-tau) were measured with Elecsys and Lumipulse assays, and Aβ1-40 with Lumipulse. CSF cut-points were defined using receiver operating characteristic (ROC) for amyloid PET positivity.
Results:
For both Elecsys and Lumipulse the p-tau/Aβ1- 42, Aβ1- 42/Aβ1- 40, and t-tau/Aβ1- 42 ratios showed similarly good concordance with amyloid PET (Elecsys: 93,90,90%; Lumipulse: 94,92,90%) and were higher than Aβ1- 42 alone (Elecsys 85%; Lumipulse 84%).
Discussion:
Biomarker ratios p-tau/Aβ1- 42, Aβ1- 42/Aβ1- 40, t-tau/Aβ1- 42 on two automated platforms show similar optimal concordance with amyloid PET in a memory clinic cohort.