Strategies for Optimization of Cryogenic Electron Tomography Data Acquisition
Published on: March 19, 2021
4.7K
Cryo-EM performance testing of hardware and data acquisition strategies
Radostin Danev1, Haruaki Yanagisawa, Masahide Kikkawa1
1Department of Cell Biology and Anatomy, Graduate School of Medicine, The University of Tokyo, 7-3-1 Hongo, Bunkyo, Tokyo 113-0033, Japan.
Microscopy (Oxford, England)
|May 10, 2021
Summary
Testing cryo-electron microscopy (cryo-EM) workflows with apoferritin ensures optimal performance. This study validates new hardware and data collection strategies for reliable cryo-EM imaging.
Area of Science:
- Structural Biology
- Biophysics
- Microscopy
Background:
- Cryo-electron microscopy (cryo-EM) adoption is rapidly increasing globally.
- New cryo-EM installations require rigorous testing to ensure optimal performance and productivity.
- Limited user experience can impact confidence in equipment capabilities.
Purpose of the Study:
- To present a standardized testing methodology for new cryo-electron microscopy (cryo-EM) hardware and data acquisition strategies.
- To evaluate the performance of a new Krios G4 electron microscope with Gatan K3 and Falcon 4 direct electron detectors.
- To assess different data acquisition strategies and the impact of environmental factors on data quality.
Main Methods:
- Utilized apoferritin as a standard test sample for initial cryo-EM workflow validation.
- Tested a Krios G4 electron microscope equipped with Gatan K3 and Falcon 4 direct electron detector cameras.
- Evaluated three beam-image shift-based data acquisition strategies, including single- and multi-shot per-hole schemes.
Main Results:
- Both Gatan K3 and Falcon 4 cameras demonstrated comparable performance.
- Single- and multi-shot per-hole acquisition strategies yielded similar results.
- Confirmed the effectiveness of Relion's software aberration correction in managing severe coma aberration.
Conclusions:
- Apoferritin is a suitable test sample for validating cryo-EM workflows and new hardware.
- The tested Krios G4 system and acquisition strategies perform reliably.
- This work provides a framework for cryo-EM teams to test and troubleshoot their equipment and data collection methods.