RENEB Inter-Laboratory comparison 2017: limits and pitfalls of ILCs

Eric Gregoire1, Joan Francesc Barquinero2, Gaetan Gruel1

  • 1Institut de Radioprotection et de Sûreté Nucléaire, Fontenay-aux-Roses, France.

Summary

This study shows that worldwide distribution of blood samples for biological dosimetry is feasible. However, current statistical tools need improvement for large inter-laboratory comparisons to accurately assess laboratory performance in radiological events.

Related Concept Videos

Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences01:20

Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences

Inductively coupled plasma–mass spectrometry (ICP–MS) is a highly selective and sensitive technique for accurate elemental analysis. Though the analysis of ICP–MS mass spectra is comparatively straightforward, it is affected by spectroscopic and non-spectroscopic interferences. Spectroscopic interferences arise when the plasma contains ionic species with an m/z value the same as the analyte ion. Spectroscopic interference can be categorized as isobaric, polyatomic ions, and...
801
Uncertainty in Measurement: Reading Instruments02:46

Uncertainty in Measurement: Reading Instruments

Counting is the type of measurement that is free from uncertainty, provided the number of objects being counted does not change during the process. Such measurements result in exact numbers. By counting the eggs in a carton, for instance, one can determine exactly how many eggs are there in the carton. Similarly, the numbers of defined quantities are also exact. For example, 1 foot is exactly 12 inches, 1 inch is exactly 2.54 centimeters, and 1 gram is exactly 0.001 kilograms. Quantities...
48.9K
Difference from Background: Limit of Detection01:05

Difference from Background: Limit of Detection

The limit of detection (LOD) is the smallest amount of analyte that can be distinguished from the background noise. The LOD value corresponds to the concentration at which the analyte signal is three times larger than the standard deviation of the blank signal. Below this value, the analyte signal cannot be differentiated from the background noise. It is calculated by dividing the calibration slope by 3 times the standard deviation of the blank signals.
The LOD indicates the presence or absence...
7.5K
Systematic Error: Methodological and Sampling Errors01:15

Systematic Error: Methodological and Sampling Errors

In the case of systematic errors, the sources can be identified, and the errors can be subsequently minimized by addressing these sources. According to the source, systematic errors can be divided into sampling, instrumental, methodological, and personal errors.
Sampling errors originate from improper sampling methods or the wrong sample population. These errors can be minimized by refining the sampling strategy. Defective instruments or faulty calibrations are the sources of instrumental...
6.0K