The influence of patient size on the overall uncertainty in radiographic dose audit
D G Sutton1, M Worrall1, K Sexton2
1Department Medical Physics, Ninewells Hospital, Dundee DD1 9SY, United Kingdom.
Abstract:
The aim of this work was to investigate the effect of patient and cohort size on the overall uncertainty associated with dose audit using radiography of the abdomen as the exemplar. Water equivalent diameterDwwas used as the surrogate for patient size and its distribution (σ(Dw)) was used to quantify the effect of sample size. The more precise the kerma area product calibration, the more patients are required in the cohort to have the same impact on the overall uncertainty. Patient sample sizes of 300-400 will result in expanded uncertainties approaching the theoretical limit of double the measurement uncertainty when audits are performed with instruments having measurement uncertainties equal to ±7%, ±10% or ±12.5%. By way of example, for a field instrument with a measurement uncertainty of ±10%, a minimum sample size of 350 is required to achieve a total expanded uncertainty of ±21%. In the case of instruments with associated measurement uncertainty of ±3.5%, patient sample sizes of 300-400 will result in expanded uncertainties of approximately ±10%. From review of the literature and comparison with the results obtained here, it is conjectured that for radiographic dose audits of all parts of the trunk the contribution to overall uncertainty due to patient and sample size could be predicted using an indicative value forσ(Dw) of 3.4 where local data is not available.
More Related Videos
10:33Expedited Radiation Biodosimetry by Automated Dicentric Chromosome Identification ADCI and Dose Estimation
Published on: September 4, 2017
06:33Construction of a Preclinical Multimodality Phantom Using Tissue-mimicking Materials for Quality Assurance in Tumor Size Measurement
Published on: July 29, 2013
Related Concept Videos
Drug Dosing: Obese Patients
Dose Size and Dosing Frequency: Determination Methods
Radiological Investigation I: X-ray and CT
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Dosage Regimen: Individualization
Propagation of Uncertainty from Random Error