Electromagnetic interference from left ventricular assist device in patients with transvenous implantable
Banveet Kaur Khetarpal1, Justin Z Lee2, Awad I Javaid1
1Internal Medicine Department, University of Nevada, Las Vegas, Nevada, USA.
Background:
Many advanced heart failure patients have both a left ventricular assist device (LVAD) and an implantable cardioverter-defibrillator (ICD). This study examines incidence, clinical impact, and management of LVAD-related EMI.
Methods:
We performed a three-center retrospective analysis of transvenous ICD implanted patients with LVAD implanted between January 1, 2005 and December 31, 2020. The primary outcome was EMI after LVAD implantation, categorized as LVAD-related noise or telemetry interference.
Results:
The rate of LVAD-related EMI among the 737 patients (mean age 58.6 ± 12.8 years) studied was 5.0%. Telemetry interference (1.5%) compromised ICD interrogation in all patients. This was resolved successfully with use of a metal shield, encased wand, radiofrequency tower, different ICD programmer or by increasing distance between ICD programmer and LVAD (n = 6). ICD replacement was required to reestablish successful communication in three patients. LVAD-related noise (3.5%) led to oversensing (n = 4), inappropriate mode switches (n = 4), noise reversion (n = 3), inhibition of pacing (n = 2), inappropriate detection as atrial fibrillation (AF) (n = 2) and inappropriate detection as ventricular tachycardia (VT) and/or ventricular fibrillation (VF) (n = 2). This noise interference persisted (n = 3), resolved spontaneously (n = 16), resolved with programming change (n = 6) or required lead revision (n = 1).
Conclusions:
EMI from LVAD impacts ICD function, although, the incidence rate is low. Physicians implanting both, LVAD in patients with ICD (more common) or ICD in patients with LVAD, should be aware of possible interferences. Telemetry failure not resolved by metal shielding was overcome by ICD generator replacement to a different manufacturer. In most cases, LVAD-related noise resolves spontaneously.
More Related Videos
Related Concept Videos
Cardiomyopathy V: Interprofessional Care
Cardiac Catheterization III: Left Heart Catheterization
Cardiac Catheterization II: Right Heart Catheterization
Cardiomyopathy II: Dilated Cardiomyopathy
Cardiac Catheterization IV: Nursing Management
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias