Reliability of Performing Multiparametric Ultrasound in Adult Livers
Jing Gao1, Rebecca Lee1, Michael Trujillo1
1Rocky Vista University, Ivins, Utah, USA.
Multiparametric ultrasound shows moderate to excellent reliability for liver shear wave velocity, attenuation, and variance. However, shear wave dispersion slope and liver-to-kidney ratio measurements need standardization for improved reproducibility in liver steatosis assessment.
Area of Science:
- Medical Imaging
- Hepatology
- Ultrasound Technology
Background:
- Multiparametric ultrasound (MPU) offers a non-invasive method for liver assessment.
- Evaluating the reliability of MPU parameters is crucial for clinical application in diagnosing liver conditions like steatosis.
Purpose of the Study:
- To assess the inter-observer and intra-observer reliability of various MPU parameters in adult livers.
- To determine the reproducibility of measurements for shear wave velocity (SWV), shear wave dispersion slope (SWD), attenuation coefficient (ATI), normalized local variance (NLV), and liver-to-kidney ratio (L/K).
Main Methods:
- 21 adult participants underwent MPU and liver magnetic resonance imaging-proton density fat fraction (MRI-PDFF) for steatosis quantification.
- Measurements included SWV, SWD, ATI, NLV, and L/K ratio.
- Intraclass correlation coefficients (ICCs) and Bland-Altman limits of agreement (LOA) were used to analyze reproducibility.
Main Results:
- ICCs for intra- and inter-observer reproducibility of most MPU parameters were above 0.75.
- SWV, ATI, and NLV demonstrated moderate to excellent repeatability and reproducibility.
- SWD and L/K ratio showed poor reliability, with wider LOA.
- Significant differences in SWV, SWD, ATI, NLV, and L/K ratio were observed between normal and steatotic livers (p < .05).
Conclusions:
- MPU parameters SWV, ATI, and NLV show promising reliability for liver assessment.
- SWD and L/K ratio measurements require further standardization to improve consistency.
- Standardized machine settings, scanning protocols, and operator training are recommended for reliable MPU of the liver.
More Related Videos
05:56Implementation of Non-invasive Point of Care Transient Elastography for Evaluation of Liver Disease in Pediatric Populations with Cystic Fibrosis
Published on: August 29, 2025
07:12Author Spotlight: Using Point-of-Care Ultrasound for Comprehensive Evaluation of the Abdominal Aorta
Published on: September 8, 2023
Related Concept Videos
Ultrasound II: Endoscopic Ultrasound and FibroScan
Endoscopic Ultrasound (EUS):
Ultrasound I: Abdominal Ultrasonography
Abdominal ultrasonography, commonly known as abdominal ultrasound, is a vital, non-invasive medical imaging technique widely used in healthcare.
Procedure:
This diagnostic tool allows the clinician to visually inspect internal structures within the abdomen, including vital organs such as the liver, gallbladder, pancreas, kidneys, and spleen.
The abdominal ultrasound process begins with applying a special gel to the patient's skin over the abdomen. This gel enhances the...
Ultrasonography
During an ultrasonography procedure, a handheld device called...
Assessment of the Abdomen II: Percussion
Percussion
Percussion is an essential...