Related Experiment Video
Updated: Nov 5, 2025

A Hydrogen-Deuterium Exchange Mass Spectrometry HDX-MS Platform for Investigating Peptide Biosynthetic Enzymes
Published on: May 4, 2020
Rapid Assessment of Pepsin Column Activity for Reliable HDX-MS Studies
Clint Vorauer1, Michael S Wrigley2, Juan P Rincon Pabon2
1Department of Medicinal Chemistry, University of Washington, Seattle, Washington 98195, United States.
Abstract:
Hydrogen/deuterium exchange with mass spectrometry (HDX-MS) is a widely used technique to probe protein structural dynamics, track conformational changes, and map protein-protein interactions. Most HDX-MS studies employ a bottom-up approach utilizing the acid active protease pepsin to digest the protein of interest, often utilizing immobilized protease in a column format. The extent of proteolytic cleavage will greatly influence data quality and presents a major source of variation in HDX-MS studies. Here, we present a simple cocktail of commonly available peptides that are substrates of pepsin and can serve as a rapid check of pepsin column activity. The peptide-based assay requires no system modifications and provides an immediate readout to check and benchmark pepsin activity across different HDX-MS platforms.