Related Experiment Video
Updated: Nov 5, 2025

Simultaneous Laryngopharyngeal and Conventional Esophageal pH Monitoring
Published on: December 14, 2020
Utility of a new automated diagnostic program in high-resolution esophageal manometry
Shiko Kuribayashi1, Junichi Akiyama2, Haruo Ikeda3
1Department of Gastroenterology and Hepatology, Gunma University Graduate School of Medicine, 3-39-15 Showa-machi, Maebashi, Gunma, 371-8511, Japan. shikokuri@yahoo.co.jp.
Background:
A new automated diagnostic program for high-resolution esophageal manometry (HREM) has been developed. This diagnostic program could detect locations of landmarks and could make final diagnoses automatically. However, the accuracy of the program is not known. The aim of this study was to evaluate the accuracy of the automated diagnostic program for HREM.
Methods:
A total of 445 studies were enrolled. An HREM system (Starlet®) was used, and esophageal motility was diagnosed using the Chicago classification v3.0. First, the locations of the upper esophageal sphincter, transition zone, lower esophageal sphincter, esophago-gastric junction, crural diaphragm and stomach were determined, and each swallow was checked manually. Then, the parameters of the Chicago classification were calculated using an analytic program of the Starlet, and diagnoses were made by three experts. Second, all study raw data were analyzed again by the automated diagnostic program. Diagnoses made by the program were compared to those made by experts to evaluate the accuracy of the diagnoses.
Results:
The new diagnostic program could identify the landmarks of each swallow, calculate the parameters and make a final diagnosis within 10 s. The diagnoses made by the automated diagnostic program were not matched to those made by experts in only 10 studies, and the overall accuracy of the new automated diagnostic program thus reached 97.8% (435/445).
Conclusions:
The new automated diagnostic program for HREM is clinically useful in terms of high diagnostic accuracy and time-saving.
More Related Videos
11:50High-frequency High-resolution Echocardiography: First Evidence on Non-invasive Repeated Measure of Myocardial Strain, Contractility, and Mitral Regurgitation in the Ischemia-reperfused Murine Heart
Published on: July 9, 2010
06:08Author Spotlight: Studying Neuromuscular Responses and Motor Neuron Plasticity in Neurodegenerative Diseases
Published on: April 19, 2024
Related Concept Videos
Ultrasound II: Endoscopic Ultrasound and FibroScan
Endoscopic Ultrasound (EUS):
Endoscopic Procedures I: Esophagogastroduodenoscopy
During an EGD, the endoscope can be used to: