Comparative analysis of in vitro accelerated glistening formation in foldable hydrophobic intraocular lenses
Tamer Tandogan1, Gerd U Auffarth2, Chul Young Choi3
1The David J. Apple International Laboratory for Ocular Pathology and International Vision Correction Research Centre (IVCRC), Department of Ophthalmology, University of Heidelberg, INF 400, 69120, Heidelberg, Germany.
Purpose:
To analyse and compare the propensity to form glistenings in 4 different types of hydrophobic acrylic intraocular lenses (IOLs): Alcon AcrySof ® MA60AC, HOYA iSert® PC-60AD, Bausch&Lomb enVista, and Kowa Avansee™ PU6A.
Methods:
We used an accelerated laboratory method to create glistenings. IOLs were first immersed in saline at 45 °C for 24 h and then at 37 °C for 2.5 h. Microvacuole (MV) density and size were documented and calculated using an image analysis program.
Results:
Median density of glistenings [MV/mm2] for Alcon AcrySof ® MA60AC was 623 (range 507-804), for HOYA iSert® PC-60AD 1358 (range 684-2699), for Bausch&Lomb enVista 2 (range 1-2), and for Kowa Avansee™ PU6A 1 (range 1-4). The prevailing MV size was: 0-5 µm for Hoya IOLs, 5-10 µm for Alcon IOLs, 20-50 µm for Bausch&Lomb IOLs, and 5-50 µm for Kowa IOLs.
Conclusions:
Glistenings could be induced in all studied IOLs using the accelerated laboratory method. The Alcon AcrySof ® MA60AC and HOYA iSert® PC-60AD IOLs showed MV of high density, while the glistenings in the Hoya IOLs were smaller in size compared to the Alcon IOLs. The MV density was minimal in the Bausch&Lomb enVista and Kowa Avansee™ PU6A IOLs. The propensity of the Alcon AcrySof ® MA60AC IOLs to form glistenings in vitro correlated with the findings of clinical results that are already published.