Displacement between anterior chamber width obtained by swept-source anterior segment optical coherence tomography
Teerajet Taechameekietichai1, Anwell Nguyen2, Sunee Chansangpetch2,3
1Faculty of Medicine, Chulalongkorn University, Bangkok, Thailand.
Purpose:
To determine the relationship between the external limbal location, represented by white-to-white (WTW) distance, and the actual angle location, represented by spur-to-spur (STS) and angle-to-angle (ATA) distances.
Methods:
166 eyes from 166 participants were imaged using CASIA2 anterior chamber optical coherence tomography (AS-OCT) and LenStar LS 900 optical biometer. The horizontal ATA and STS were measured using the swept-source Fourier-domain AS-OCT (CASIA2). The horizontal WTW was automatically measured using LenStar. The displacement lengths (DL) between WTW-STS and WTW-ATA were calculated. Bland-Altman plots and intraclass correlation were performed.
Results:
The study showed that WTW has a positive correlation with STS (ICC = 0.82, p<0.001) and ATA (ICC = 0.82, p<0.001). The Bland-Altman analysis demonstrated that the mean difference of WTW-STS is 0.10 mm (95% CI 0.06 to 0.14 mm) with limits of agreement of -0.42 to 0.63 mm between WTW and STS, and the mean difference of WTW-ATA is 0.10 mm (95% CI 0.06 to 0.15 mm) with limits of agreement of -0.48 to 0.64 mm between WTW and ATA. Linear regression with adjustment showed that a WTW value greater than 12.07 mm is associated with a greater DL (WTW-STS DL ß 0.18, p = 0.003; WTW-ATA DL ß 0.14, p = 0.03).
Conclusions:
Greater WTW was significantly associated with higher displacement of WTW from the two distances representing anterior chamber width. External limbal location may not accurately represent the actual angle location in eyes with larger WTW.