Purpose:
Race-adjusted interpretation of data from Cirrus high-definition OCT (HD-OCT) devices is not standard practice. The aim of this study is to evaluate differences in peripapillary retinal nerve fiber layer (RNFL) thickness between healthy Black Americans and the Cirrus HD-OCT normative database.
Design:
This is a cross-sectional observational study using control patients recruited from the greater Philadelphia, Pennsylvania, area.
Participants:
A total of 466 eyes were included in this study. Subjects were retrospectively identified from the control cohort of the Primary Open-Angle African American Glaucoma Genetics (POAAGG) study.
Methods:
Using an age-stratified or linear regression method, we reclassified white-green-yellow-red color probability codes for RNFL thicknesses by quadrant.
Main Outcome Measures:
The distribution of reclassified color codes was compared with the expected 5%-90%-4%-1% percentiles and to the original color codes by the Cirrus machine.
Results:
Average RNFL thickness in the POAAGG control cohort was thinner than in the Cirrus normative database in all except the nasal quadrant. The original color codes of the POAAGG cohort did not fall into the expected distributions, with more RNFL measurements assigned as white and red codes than expected (9.5% and 1.7%) and fewer measurements assigned as green and yellow codes than expected (85.3% and 3.5%) (P < 0.001). Compared with the original Cirrus machine, reclassification using linear regression produced color codes closest to the expected distributions (P = 0.09). The proportion of abnormal results shifted closer to the expected 5% in the nasal (1.3%, P < 0.001 vs. 3.0%, P = 0.048) and temporal (8.2%, P = 0.002 vs. 3.6%, P = 0.18) quadrants.
Conclusions:
Results further establish the presence of structural differences in the RNFL of Black American patients. Color code reclassification suggests that the existing Cirrus database may not be accurately evaluating glaucomatous nerves in patients of African descent. This study addresses an unmet need to assess Cirrus HD-OCT color probability codes in a Black American population.
Related Concept Videos
Expected Frequencies in Goodness-of-Fit Tests
Probability Histograms
Sign Test for Median of Single Population
Testing a Claim about Population Proportion
There are two methods of testing a claim about a population proportion: (1) Using the sample proportion from the data where a binomial distribution is approximated to the normal distribution and (2) Using the binomial probabilities calculated from the data.
The first method uses normal distribution as an approximation to the binomial distribution. The requirements are as follows: sample size is large...
Hazard Ratio
For example, in a clinical trial...
Critical Region, Critical Values and Significance Level
In hypothesis testing, a sample statistic is converted to a test statistic using z, t, or chi-square distribution. A critical region is an area under the curve in probability distributions demarcated by the critical value. When the test statistic falls in this region, it suggests that the null hypothesis must be rejected. As this region contains all those values of the...