Ray-tracing Calculation Using Scheimpflug Tomography of Diffractive Extended Depth of Focus IOLs Following Myopic
Purpose:
To evaluate a ray-tracing formula for intraocular lens (IOL) calculation of diffractive extended depth of focus IOLs after myopic laser in situ keratomileusis (LASIK) compared to formulas from an established online calculator.
Methods:
This retrospective, consecutive case series included patients after cataract surgery with implantation of an extended depth of focus (EDOF) IOL (AT LARA, Carl Zeiss Meditec; Symfony, Johnson & Johnson) and a history of myopic LASIK. Preoperative assessments included biometry (IOLMaster; Carl Zeiss Meditec) and corneal tomography, including true net power (TNP) (Pentacam; Oculus Optikgeräte GmbH). To evaluate the measurements, the simulated keratometry values (SimK) were compared to the TNP. Regarding IOL calculation, the mean prediction error, mean and median absolute prediction error (MAE and MedAE), and number of eyes within ±0.50, ±1.00, and ±2.00 diopters (D) from the Haigis-L, Shammas, and Barrett True K No History formulas to the Potvin-Hill and Haigis with TNP (Pentacam) formulas were compared.
Results:
Thirty-six eyes matched the inclusion criteria with a mean spherical equivalent of -6.26 ± 3.25 diopters (D) preoperatively and -0.79 ± 0.75 D postoperatively. The mean difference from SimK and TNP was significantly different from zero (P < .001; -1.24 ± 0.81 D). The best performing formulas by MedAE were the Potvin-Hill and Barrett True K No History (0.39 ± 0.78 and 0.64 ± 1.00 D). The formula with the most eyes within ±0.50 D was the Potvin-Hill (64%), followed by the Barrett True K No History (44%). For MAE and percentage of eyes within ±0.50 D, the Potvin-Hill formula was significantly better than the Haigis-L, Shammas, and Haigis-TNP formulas (P < .05).
Conclusions:
Calculation of IOLs in patients who had LASIK remains less predicable than calculations for virgin eyes. Using ray-tracing to calculate diffractive EDOF IOLs after myopic LASIK, the Potvin-Hill formula outperformed established formulas in terms of the percentage within target refraction and the MAE. [.
More Related Videos
05:46Author Spotlight: Advancements in Refractive Surgical Correction for Presbyopia and Exploring Postoperative Visual Acuity
Published on: September 20, 2024
05:14Comparison of Agreement and Accuracy using Binocular Wavefront Optometer with Autorefractor and Phoropter
Published on: September 16, 2025